Determination of Cefradin by RP-HPLC
Jiang Jian
Abstract
Jiang Jian
Abstract
A RP-HPLC method for the determination of cefradin in raw drug was established.A column of YQG-ODS was used with a mobile phase of V(CH3OH)∶V(4%HAc-NaAc)=3∶7 solution.The UV detector was set at 262 nm.The external standard method was used.There was a good linear correlation in the range of 0.027~0.29 g·L-1,Y(μV·s)=1.88×107X(g·L-1)+20 876,r=0.998 3.The average recovery was 99.7%,RSD was 0.57%(n=3).The elapsed time of a chromatographic analysis was 12 min.The method is rapid,accurate and reliable.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
A RP-HPLC method for the determination of cefradin in raw drug was established.A column of YQG-ODS was used with a mobile phase of V(CH3OH)∶V(4%HAc-NaAc)=3∶7 solution.The UV detector was set at 262 nm.The external standard method was used.There was a good linear correlation in the range of 0.027~0.29 g·L-1,Y(μV·s)=1.88×107X(g·L-1)+20 876,r=0.998 3.The average recovery was 99.7%,RSD was 0.57%(n=3).The elapsed time of a chromatographic analysis was 12 min.The method is rapid,accurate and reliable.
Key concepts: Chromatography, High-performance liquid chromatography, Chemistry, Linear correlation, Linear range, Analytical Chemistry (journal), Materials science, Detection limit