Evaluation on alarm prompt function of Sysmex XE-2100 analyzer in detecting abnormal cells of peripheral blood in children
Han Jing-jin
Abstract
Han Jing-jin
Abstract
Objective To appraise the alarm prompt function of the Sysmex XE-2100 analyzer to detect abnormal cells in peripheral blood of children.Methods To appraise the reliability of alarm prompt function of the Sysmex XE-2100 analyzer for analyzing abnormal cells in peripheral blood of children,804 samples were examined repeatedly by he Sysmex XE-2100 analyzer and microscope measurement,and a statistical comparison was also made.Results The sensitivity and specificity of the alarming system for abnormal cells were 100% and 69.80%,respectively.The positive predictive value was 54.10%,and the negative predictive value was 100%.The efficiency rate was 77.74%,the false positive rate was 45.90%.Conclusion The Sysmex XE-2100 analyzer has a good sensitivity in analyzing abnormal cells of peripheral blood of children,but its false positive rate is also high,so those samples with abnormal alarm information need to be ascertained by smear for microscope test.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To appraise the alarm prompt function of the Sysmex XE-2100 analyzer to detect abnormal cells in peripheral blood of children.Methods To appraise the reliability of alarm prompt function of the Sysmex XE-2100 analyzer for analyzing abnormal cells in peripheral blood of children,804 samples were examined repeatedly by he Sysmex XE-2100 analyzer and microscope measurement,and a statistical comparison was also made.Results The sensitivity and specificity of the alarming system for abnormal cells were 100% and 69.80%,respectively.The positive predictive value was 54.10%,and the negative predictive value was 100%.The efficiency rate was 77.74%,the false positive rate was 45.90%.Conclusion The Sysmex XE-2100 analyzer has a good sensitivity in analyzing abnormal cells of peripheral blood of children,but its false positive rate is also high,so those samples with abnormal alarm information need to be ascertained by smear for microscope test.
Key concepts: Peripheral blood, Spectrum analyzer, Medicine, False positive rate, ALARM, Pathology, Predictive value, Immunology