Analysis and solutions of no platelet-associated parameters displayed by XS-1000i hematology analyzer
Can Lu
Abstract
Can Lu
Abstract
Objective To investigate the cause and solution of no platelet-associated parameters displayed by the Sysmex XS-1000i hematology analyzer.Methods Swear,Wright Stain,then microscopic examination the red blood cells and platelets abnormal morphology in 123 cases of no platelet-associated parameters were analyzed by the Sysmex XS-1000i hematology analyzer.Results In 123 cases of no platelet-associated parameters analyzed by the Sysmex XS-1000i hematology analyzer,microscopic examination showed varying degrees of abnormal morphology of red blood cells and(or) abnormal morphology thrombocytosis.Conclusion Encountered the lack of platelet-associated parameters analyzed by the Sysmex XS-1000i hematology analyzer must use the method of Wright stain and microscopic examination as a supplement,which can provide a detailed description of the test results.Platelet can be detected using reference method in laboratory conditions.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To investigate the cause and solution of no platelet-associated parameters displayed by the Sysmex XS-1000i hematology analyzer.Methods Swear,Wright Stain,then microscopic examination the red blood cells and platelets abnormal morphology in 123 cases of no platelet-associated parameters were analyzed by the Sysmex XS-1000i hematology analyzer.Results In 123 cases of no platelet-associated parameters analyzed by the Sysmex XS-1000i hematology analyzer,microscopic examination showed varying degrees of abnormal morphology of red blood cells and(or) abnormal morphology thrombocytosis.Conclusion Encountered the lack of platelet-associated parameters analyzed by the Sysmex XS-1000i hematology analyzer must use the method of Wright stain and microscopic examination as a supplement,which can provide a detailed description of the test results.Platelet can be detected using reference method in laboratory conditions.
Key concepts: Hematology analyzer, Platelet, Hematology, Thrombocytosis, Medicine, Stain, Pathology, Internal medicine