Evaluation of detection performance of DXI800 chemiluminescence analyzer according to EP10-A2 document of CLSI
Yang Fang-hu
Abstract
Yang Fang-hu
Abstract
Objective To evaluate the detection performance of DXI800chemiluminescence analyzer(DXI800 analyzer)according to EP10-A2document of Clinic and Laboraroty Standard Institute(CLSI).Methods According to EP10-A2document,human growth hormone(hGH)in quality control serum were continuously detected for five days with a certain order,then the deviation,total imprecision,intercept,slope rate,non-linearity,cross-contamination and drift were analyzed by using Microsoft Excel software.Results The linearity between measured value and target value was relatively well(Y=0.992 X+0.083,r2=0.978)when the concentrations of hGH were 2.15,5.47 and 10.80mmol/L,with bias of 0.066,0.265and 0.055 mmol/L,and total imprecision of 6.57%,8.53% and 4.56%,respectively.Simple sample t test indicated that the difference of intercept,slope rate,non-linearity,crosscontamination and drift was no statistically significant(P0.05).Conclusion The deviation,total imprecision,intercept,slope rate,non-linearity,cross-contamination and drift of DXI800analyzer might be within permission limit,and could satisfy the request of clinical application and scientific research of hGH.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To evaluate the detection performance of DXI800chemiluminescence analyzer(DXI800 analyzer)according to EP10-A2document of Clinic and Laboraroty Standard Institute(CLSI).Methods According to EP10-A2document,human growth hormone(hGH)in quality control serum were continuously detected for five days with a certain order,then the deviation,total imprecision,intercept,slope rate,non-linearity,cross-contamination and drift were analyzed by using Microsoft Excel software.Results The linearity between measured value and target value was relatively well(Y=0.992 X+0.083,r2=0.978)when the concentrations of hGH were 2.15,5.47 and 10.80mmol/L,with bias of 0.066,0.265and 0.055 mmol/L,and total imprecision of 6.57%,8.53% and 4.56%,respectively.Simple sample t test indicated that the difference of intercept,slope rate,non-linearity,crosscontamination and drift was no statistically significant(P0.05).Conclusion The deviation,total imprecision,intercept,slope rate,non-linearity,cross-contamination and drift of DXI800analyzer might be within permission limit,and could satisfy the request of clinical application and scientific research of hGH.
Key concepts: Spectrum analyzer, Linearity, Chemiluminescence, Detection limit, Mathematics, Microsoft excel, Statistics, Recovery rate