Preliminary evaluation of quantitative laboratory method for estradiol(E2) in DXI800 photometric analyzer with EP10-A
Qinshi Pan
Abstract
Qinshi Pan
Abstract
Objective:To preliminary evaluation of quantitative laboratory method of DXI800 photometric analyzer with EP10-A.Methods: According to the NCCLS Document EP10-A and Chinese public health profession standard Preliminary Assessment of Quantitation Clinic Analysis Method(WS/T228-2002),collected blood sample of patients and detected E2 continuously for five days with a certain order in the DXI800 analyzer,then calculated deviation,total imprecision,intercept,slope rate,nonlinearity,cross-contamination,drift and carried out statistical analysis.Results: It showed that bias are 1.34 pmo/L,-7.90 pmo/L,9.53 pmo/L,total imprecision are 5.15%,3.79%,7.36% and intercept,slope rate,nonlinearity,cross-contamination,drift were 19.042 pmo/L(P0.01),0.946(P0.01),-0.627(P0.01),0.12(P0.01),-2.226(P0.01)respectively.Conclusion: The deviation,total imprecision,intercept,slope rate,nonlinearity,cross-contamination,drift of the DXI800 analyzer were within permission limit,so the analyzer could satisfy the clinical application and scientific research request of E2.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective:To preliminary evaluation of quantitative laboratory method of DXI800 photometric analyzer with EP10-A.Methods: According to the NCCLS Document EP10-A and Chinese public health profession standard Preliminary Assessment of Quantitation Clinic Analysis Method(WS/T228-2002),collected blood sample of patients and detected E2 continuously for five days with a certain order in the DXI800 analyzer,then calculated deviation,total imprecision,intercept,slope rate,nonlinearity,cross-contamination,drift and carried out statistical analysis.Results: It showed that bias are 1.34 pmo/L,-7.90 pmo/L,9.53 pmo/L,total imprecision are 5.15%,3.79%,7.36% and intercept,slope rate,nonlinearity,cross-contamination,drift were 19.042 pmo/L(P0.01),0.946(P0.01),-0.627(P0.01),0.12(P0.01),-2.226(P0.01)respectively.Conclusion: The deviation,total imprecision,intercept,slope rate,nonlinearity,cross-contamination,drift of the DXI800 analyzer were within permission limit,so the analyzer could satisfy the clinical application and scientific research request of E2.
Key concepts: Spectrum analyzer, Detection limit, Recovery rate, Contamination, Analytical Chemistry (journal), Relative standard deviation, Standard deviation, Chromatography