Preliminary functional evaluation on chemiluminescence immunoassay system by CLSI EP10-A2
Zhang Da-xiu
Abstract
Zhang Da-xiu
Abstract
Objective To evaluate the performance of UniCel DXI800 automated chemiluminescence immunoassay system.Methods The indexes of E2,FSH,PRL,TT3 were detected according to the CLSI Document EP10-A2 preliminary evaluation of quantitative clinical laboratory methods:approved guideline.Results The high,medium,low total imprecision of E2 were 7.79%,5.21%,4.48%,the bias of them were-2.45%,-1.90%,-0.40%.The high,medium,low total imprecision of FSH were 3.58%,4.81%,4.71%,the bias of them were 6.84%,4.73%,0.93%.The high,medium,low total imprecision of PRL were 2.63%,3.10%,2.24%,the bias of them were-0.03%,-0.65%,-0.50%.The high,medium,low total imprecision of TT3 were 2.71%,1.60%,2.05%,the bias of them were-0.01%,0.01%,-0.05%.The indexes of slope,intercept, precision,carryover rate,linearity and total repeatability accorded with the prescriptive range.Conclusion The performance of DXI800 automated chemiluminescence immunoassay system is reliable and suitable for clinical laboratory to use.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To evaluate the performance of UniCel DXI800 automated chemiluminescence immunoassay system.Methods The indexes of E2,FSH,PRL,TT3 were detected according to the CLSI Document EP10-A2 preliminary evaluation of quantitative clinical laboratory methods:approved guideline.Results The high,medium,low total imprecision of E2 were 7.79%,5.21%,4.48%,the bias of them were-2.45%,-1.90%,-0.40%.The high,medium,low total imprecision of FSH were 3.58%,4.81%,4.71%,the bias of them were 6.84%,4.73%,0.93%.The high,medium,low total imprecision of PRL were 2.63%,3.10%,2.24%,the bias of them were-0.03%,-0.65%,-0.50%.The high,medium,low total imprecision of TT3 were 2.71%,1.60%,2.05%,the bias of them were-0.01%,0.01%,-0.05%.The indexes of slope,intercept, precision,carryover rate,linearity and total repeatability accorded with the prescriptive range.Conclusion The performance of DXI800 automated chemiluminescence immunoassay system is reliable and suitable for clinical laboratory to use.
Key concepts: Chemiluminescence, Repeatability, Immunoassay, Chromatography, Chemiluminescent immunoassay, Guideline, Medicine, Chemistry