The observation of imprecision and linearity on detection of pepsinogen II by latex-enhanced immunoturbidimetry assay
Songli Li
Abstract
Songli Li
Abstract
Objective To evaluate primarily the detection of pepsinogen Ⅱ(PGⅡ) by latex-enhanced immunoturbidimetry assay,and provide the reference for clinical application.Methods According to the National Committee for Clinical Laboratory Standards(NCCLS) EP5-A and EP10-A2 documents,the detection of PGⅡ by latex-enhanced immunoturbidimetry assay was evaluated primarily.Results The within-run,between-run,between-day and total coefficients of variation(CV) of samples with low concentration of PGⅡ were 2.05%,1.60%,3.92% and 4.70% respectively.The within-run,between-run,between-day and total CV of samples with high concentration of PGⅡ were 1.70%,1.21%,3.79% and 4.32% respectively.When PGⅡ concentrations were 9.6,24.7 and 39.8 μg/L,the relative biases were-0.18,-0.47 and-0.39 μg/L respectively.The total imprecisions were 4.98%,2.39% and 2.71% respectively.There was no statistical significance in total carry-over(P0.05).The equation of linear regression of PGⅡ was Y =0.975 7X-0.016 8,and the coefficient of determination(R2)= 0.998 7.Conclusions The latex-enhanced immunoturbidimetry assay for the PGⅡ detection with good repeatability meets the clinical application requirements of EP5-A document for precision and the requirements of EP10-A2 document for the linearity,bias,total imprecision and carry-over.It is suitable for the clinical application.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To evaluate primarily the detection of pepsinogen Ⅱ(PGⅡ) by latex-enhanced immunoturbidimetry assay,and provide the reference for clinical application.Methods According to the National Committee for Clinical Laboratory Standards(NCCLS) EP5-A and EP10-A2 documents,the detection of PGⅡ by latex-enhanced immunoturbidimetry assay was evaluated primarily.Results The within-run,between-run,between-day and total coefficients of variation(CV) of samples with low concentration of PGⅡ were 2.05%,1.60%,3.92% and 4.70% respectively.The within-run,between-run,between-day and total CV of samples with high concentration of PGⅡ were 1.70%,1.21%,3.79% and 4.32% respectively.When PGⅡ concentrations were 9.6,24.7 and 39.8 μg/L,the relative biases were-0.18,-0.47 and-0.39 μg/L respectively.The total imprecisions were 4.98%,2.39% and 2.71% respectively.There was no statistical significance in total carry-over(P0.05).The equation of linear regression of PGⅡ was Y =0.975 7X-0.016 8,and the coefficient of determination(R2)= 0.998 7.Conclusions The latex-enhanced immunoturbidimetry assay for the PGⅡ detection with good repeatability meets the clinical application requirements of EP5-A document for precision and the requirements of EP10-A2 document for the linearity,bias,total imprecision and carry-over.It is suitable for the clinical application.
Key concepts: Repeatability, Coefficient of variation, Chromatography, Linear regression, Detection limit, Coefficient of determination, Medicine, Chemistry