Determination of Baicalin in Chaigejieji Decoction by HPLC
XU Yi-liang
Abstract
XU Yi-liang
Abstract
Objective To establish a HPLC method for determining Baicalin in Chaigejieji decoction. Method Baicalin was separated on ODS Kromasil C18 column with a mobile phase of methanol-water- phosphoric acid (47∶53∶0.2) and detected at 280 nm with a flow rate of 1.0 mL/min. Results The calibration curve was linear at the range of 0.3373~1.6865 μg (r =0.9997). The average recovery of Baicalin was 97.87% and RSD was 1.66%. Conclusion It is a simple and accurate method for the determination of Baicalin in Chaigejieji decoction.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To establish a HPLC method for determining Baicalin in Chaigejieji decoction. Method Baicalin was separated on ODS Kromasil C18 column with a mobile phase of methanol-water- phosphoric acid (47∶53∶0.2) and detected at 280 nm with a flow rate of 1.0 mL/min. Results The calibration curve was linear at the range of 0.3373~1.6865 μg (r =0.9997). The average recovery of Baicalin was 97.87% and RSD was 1.66%. Conclusion It is a simple and accurate method for the determination of Baicalin in Chaigejieji decoction.
Key concepts: Baicalin, Decoction, Chromatography, Calibration curve, High-performance liquid chromatography, Chemistry, Phosphoric acid, Linear range