The comparison of intraocular pressure measurement between non-contact tonometer and Goldmann applanation tonometer

Ye Wen

Open publisher page 0 citations

Abstract

Objective:To compare the difference of intraocular pressure measured by non-contact tonometer(NCT)and Goldmann applanation tonometer.Methods:The intraocular pressure(IOP) of 112 persons(220 eyes) were measured with NCT and Goldmann applanation tonometer respectively, and the central corneal thickness were also measured with ultrasonic pachymeter. The refraction diopter were checked with autorefractor.Results:The IOP for NCT and Goldmann applanation tonometry were 1.98±0.69kPa and 2.34±0.77kPa respectively. The IOP with NCT was lower than that with Goldmann applanation tonometer(0.36±0.37kPa,P0.05).The IOP of NCT was positive relative to the central corneal thickness, and negative relative to the refraction diopter.Conclusion:The IOP with NCT was lower than that with Goldmann applanation tonometer. The critical IOP measured with NCT should be rechecked with Goldmann applanation tonometer.

About this research paper

What this paper is about

Objective:To compare the difference of intraocular pressure measured by non-contact tonometer(NCT)and Goldmann applanation tonometer.Methods:The intraocular pressure(IOP) of 112 persons(220 eyes) were measured with NCT and Goldmann applanation tonometer respectively, and the central corneal thickness were also measured with ultrasonic pachymeter. The refraction diopter were checked with autorefractor.Results:The IOP for NCT and Goldmann applanation tonometry were 1.98±0.69kPa and 2.34±0.77kPa respectively. The IOP with NCT was lower than that with Goldmann applanation tonometer(0.36±0.37kPa,P0.05).The IOP of NCT was positive relative to the central corneal thickness, and negative relative to the refraction diopter.Conclusion:The IOP with NCT was lower than that with Goldmann applanation tonometer. The critical IOP measured with NCT should be rechecked with Goldmann applanation tonometer.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective:To compare the difference of intraocular pressure measured by non-contact tonometer(NCT)and Goldmann applanation tonometer.Methods:The intraocular pressure(IOP) of 112 persons(220 eyes) were measured with NCT and Goldmann applanation tonometer respectively, and the central corneal thickness were also measured with ultrasonic pachymeter. The refraction diopter were checked with autorefractor.Results:The IOP for NCT and Goldmann applanation tonometry were 1.98±0.69kPa and 2.34±0.77kPa respectively. The IOP with NCT was lower than that with Goldmann applanation tonometer(0.36±0.37kPa,P0.05).The IOP of NCT was positive relative to the central corneal thickness, and negative relative to the refraction diopter.Conclusion:The IOP with NCT was lower than that with Goldmann applanation tonometer. The critical IOP measured with NCT should be rechecked with Goldmann applanation tonometer.

Key concepts: Goldmann Applanation Tonometer, Intraocular pressure, Dioptre, Medicine, Ophthalmology, Autorefractor, Applanation tonometry, Optometry

Related papers

Back to paper searchBrowse research topicsOriginal source
The comparison of intraocular pressure measurement between non-contact tonometer and Goldmann applanation tonometer — Research Paper | ScholarLens