Perfunnance evaluation on a latex enhanced immunoturbidimetricβ_2-microglobulin kit
Li Hu
Abstract
Li Hu
Abstract
Objective To evaluate the performance of a latex enhanced immunoturbidimetricβ2-microglobulin kit in clinical use. Methods The lower limit of detection,precision,accuracy and linearity range was determined on the automatic chemistry analyzer Hitachi 7170S. Results The lower limit of detection was 0. 09mg/L. The within-run precision was less than 4% while the between-run precision was less than 8%. The accuracy was 99.29 % and the linearity range was 0.2~18mg/L. Conclusion Theβ2-microgiobulin kit shows good performance in clinical use.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To evaluate the performance of a latex enhanced immunoturbidimetricβ2-microglobulin kit in clinical use. Methods The lower limit of detection,precision,accuracy and linearity range was determined on the automatic chemistry analyzer Hitachi 7170S. Results The lower limit of detection was 0. 09mg/L. The within-run precision was less than 4% while the between-run precision was less than 8%. The accuracy was 99.29 % and the linearity range was 0.2~18mg/L. Conclusion Theβ2-microgiobulin kit shows good performance in clinical use.