Comparative study and deviation evaluation the results of ALT detected by two biochemical detection systems
Xu Zhang
Abstract
Xu Zhang
Abstract
Objective To develop the contrast test and deviation evaluation to ALT results which detected by 2 biochemical detection systems,and explored the comparability of the 2 methods.Method According to EP9-A document of NCCLS,Roche INTEGRA 400 Plus detection system as control,OLIMPUS AU640 detection system was the empirical method,alanine aminotransferase(ALT) in 40 fresh sera were measured by the 2 methods,calculated the correlation coefficient,linear regression equation and expectant deviation,the results which detected by the 2 methods were evaluated according to CLIA'88,judgment the clinical acceptability of the 2 methods.Results The correlation coefficient R1 was 0.9990.975,R2 was 0.9990.95,correlation was well,the expectant deviation in 1/2 of CLIA88 allowable error.Conclusions The ALT results detected by Roche INTEGRA 400 Plus and OLIMPUS AU640 had comparability,and it can be detected by the 2 detection systems.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To develop the contrast test and deviation evaluation to ALT results which detected by 2 biochemical detection systems,and explored the comparability of the 2 methods.Method According to EP9-A document of NCCLS,Roche INTEGRA 400 Plus detection system as control,OLIMPUS AU640 detection system was the empirical method,alanine aminotransferase(ALT) in 40 fresh sera were measured by the 2 methods,calculated the correlation coefficient,linear regression equation and expectant deviation,the results which detected by the 2 methods were evaluated according to CLIA'88,judgment the clinical acceptability of the 2 methods.Results The correlation coefficient R1 was 0.9990.975,R2 was 0.9990.95,correlation was well,the expectant deviation in 1/2 of CLIA88 allowable error.Conclusions The ALT results detected by Roche INTEGRA 400 Plus and OLIMPUS AU640 had comparability,and it can be detected by the 2 detection systems.
Key concepts: Correlation coefficient, Alanine aminotransferase, Linear regression, Mathematics, Statistics, Linear correlation, Correlation, Medicine