Comparative study and deviation evaluation the results of ALT detected by two biochemical detection systems

Xu Zhang

Open publisher page 0 citations

Abstract

Objective To develop the contrast test and deviation evaluation to ALT results which detected by 2 biochemical detection systems,and explored the comparability of the 2 methods.Method According to EP9-A document of NCCLS,Roche INTEGRA 400 Plus detection system as control,OLIMPUS AU640 detection system was the empirical method,alanine aminotransferase(ALT) in 40 fresh sera were measured by the 2 methods,calculated the correlation coefficient,linear regression equation and expectant deviation,the results which detected by the 2 methods were evaluated according to CLIA'88,judgment the clinical acceptability of the 2 methods.Results The correlation coefficient R1 was 0.9990.975,R2 was 0.9990.95,correlation was well,the expectant deviation in 1/2 of CLIA88 allowable error.Conclusions The ALT results detected by Roche INTEGRA 400 Plus and OLIMPUS AU640 had comparability,and it can be detected by the 2 detection systems.

About this research paper

What this paper is about

Objective To develop the contrast test and deviation evaluation to ALT results which detected by 2 biochemical detection systems,and explored the comparability of the 2 methods.Method According to EP9-A document of NCCLS,Roche INTEGRA 400 Plus detection system as control,OLIMPUS AU640 detection system was the empirical method,alanine aminotransferase(ALT) in 40 fresh sera were measured by the 2 methods,calculated the correlation coefficient,linear regression equation and expectant deviation,the results which detected by the 2 methods were evaluated according to CLIA'88,judgment the clinical acceptability of the 2 methods.Results The correlation coefficient R1 was 0.9990.975,R2 was 0.9990.95,correlation was well,the expectant deviation in 1/2 of CLIA88 allowable error.Conclusions The ALT results detected by Roche INTEGRA 400 Plus and OLIMPUS AU640 had comparability,and it can be detected by the 2 detection systems.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To develop the contrast test and deviation evaluation to ALT results which detected by 2 biochemical detection systems,and explored the comparability of the 2 methods.Method According to EP9-A document of NCCLS,Roche INTEGRA 400 Plus detection system as control,OLIMPUS AU640 detection system was the empirical method,alanine aminotransferase(ALT) in 40 fresh sera were measured by the 2 methods,calculated the correlation coefficient,linear regression equation and expectant deviation,the results which detected by the 2 methods were evaluated according to CLIA'88,judgment the clinical acceptability of the 2 methods.Results The correlation coefficient R1 was 0.9990.975,R2 was 0.9990.95,correlation was well,the expectant deviation in 1/2 of CLIA88 allowable error.Conclusions The ALT results detected by Roche INTEGRA 400 Plus and OLIMPUS AU640 had comparability,and it can be detected by the 2 detection systems.

Key concepts: Correlation coefficient, Alanine aminotransferase, Linear regression, Mathematics, Statistics, Linear correlation, Correlation, Medicine

Related papers

Back to paper searchBrowse research topicsOriginal source
Comparative study and deviation evaluation the results of ALT detected by two biochemical detection systems — Research Paper | ScholarLens