Determination of carbamazepine concentration in human serum by HPLC

Hong Zhang

Open publisher page 0 citations

Abstract

OBJECTIVE To establish an HPLC method for determining serum level of Carbamazepine.METHODS A symmetryshield RP18 column(5μm,3.9×150mm)was used.The mobile phase,methanol-phosphate solution(55∶45) was used at a flow rate 1.0mL·min-1,The detector is Waters 2487 Dual λ Absorbance Detector,detector was set at the wavelength of 210nm.RESULTS The calibration curve was linear in the range of 2.0 to 20.0μg·mL-1.The average recovery was 96%,and the RSD within-day and between-day were all less than 5%.CONCLUSION This method is suitable for therapeutic drug monitoring of Carbamazepine.

About this research paper

What this paper is about

OBJECTIVE To establish an HPLC method for determining serum level of Carbamazepine.METHODS A symmetryshield RP18 column(5μm,3.9×150mm)was used.The mobile phase,methanol-phosphate solution(55∶45) was used at a flow rate 1.0mL·min-1,The detector is Waters 2487 Dual λ Absorbance Detector,detector was set at the wavelength of 210nm.RESULTS The calibration curve was linear in the range of 2.0 to 20.0μg·mL-1.The average recovery was 96%,and the RSD within-day and between-day were all less than 5%.CONCLUSION This method is suitable for therapeutic drug monitoring of Carbamazepine.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

OBJECTIVE To establish an HPLC method for determining serum level of Carbamazepine.METHODS A symmetryshield RP18 column(5μm,3.9×150mm)was used.The mobile phase,methanol-phosphate solution(55∶45) was used at a flow rate 1.0mL·min-1,The detector is Waters 2487 Dual λ Absorbance Detector,detector was set at the wavelength of 210nm.RESULTS The calibration curve was linear in the range of 2.0 to 20.0μg·mL-1.The average recovery was 96%,and the RSD within-day and between-day were all less than 5%.CONCLUSION This method is suitable for therapeutic drug monitoring of Carbamazepine.

Key concepts: Carbamazepine, Chromatography, Calibration curve, Chemistry, High-performance liquid chromatography, Absorbance, Phosphate buffered saline, Linear range

Related papers

Back to paper searchBrowse research topicsOriginal source
Determination of carbamazepine concentration in human serum by HPLC — Research Paper | ScholarLens