Analysis to Flag-warning of Automated Hematology Analyzer and Blood-smear Microscopy Examination

Liu Cui-yu

Open publisher page 0 citations

Abstract

Objective To discuss the relationship between Flag-warning of automated Hematology analyzer and the result of blood-smears microscopic examination.Methods 867 sample were analyzed with Beckman coulter LH-750 automated hematology analyzer.According to the flag-warning of LH-750,the abnormal samples were smeared,stained,and optical examined with microscope.We compared the results of the two methods.Results The positive rate of hematology analyzer's flag-warning is 21.3%,blood-smears microscopic examination is 9.2%.There were statistical meaning between the two methods(χ2=49.11,P0.01).But there were a little failed flag-warning for detecting immature Granulocyte,stab nucleus,viriant lymphocyte,eosinophil,monocyte,nucleated erythrocyte,etc.Conclusion The flag-warning of LH-750 automated hematology analyzer had a failed positive rate.The microscopic examination for abnormal sample based on flag-warning of hematology analyzer,can make sure to check out the positive samples,it can also correct the failed positive results.So we say morphological examination is a standard method.

About this research paper

What this paper is about

Objective To discuss the relationship between Flag-warning of automated Hematology analyzer and the result of blood-smears microscopic examination.Methods 867 sample were analyzed with Beckman coulter LH-750 automated hematology analyzer.According to the flag-warning of LH-750,the abnormal samples were smeared,stained,and optical examined with microscope.We compared the results of the two methods.Results The positive rate of hematology analyzer's flag-warning is 21.3%,blood-smears microscopic examination is 9.2%.There were statistical meaning between the two methods(χ2=49.11,P0.01).But there were a little failed flag-warning for detecting immature Granulocyte,stab nucleus,viriant lymphocyte,eosinophil,monocyte,nucleated erythrocyte,etc.Conclusion The flag-warning of LH-750 automated hematology analyzer had a failed positive rate.The microscopic examination for abnormal sample based on flag-warning of hematology analyzer,can make sure to check out the positive samples,it can also correct the failed positive results.So we say morphological examination is a standard method.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To discuss the relationship between Flag-warning of automated Hematology analyzer and the result of blood-smears microscopic examination.Methods 867 sample were analyzed with Beckman coulter LH-750 automated hematology analyzer.According to the flag-warning of LH-750,the abnormal samples were smeared,stained,and optical examined with microscope.We compared the results of the two methods.Results The positive rate of hematology analyzer's flag-warning is 21.3%,blood-smears microscopic examination is 9.2%.There were statistical meaning between the two methods(χ2=49.11,P0.01).But there were a little failed flag-warning for detecting immature Granulocyte,stab nucleus,viriant lymphocyte,eosinophil,monocyte,nucleated erythrocyte,etc.Conclusion The flag-warning of LH-750 automated hematology analyzer had a failed positive rate.The microscopic examination for abnormal sample based on flag-warning of hematology analyzer,can make sure to check out the positive samples,it can also correct the failed positive results.So we say morphological examination is a standard method.

Key concepts: Hematology analyzer, Medicine, Flag (linear algebra), Spectrum analyzer, Hematology, Atypical Lymphocyte, Blood smear, False positive rate

Related papers

Back to paper searchBrowse research topicsOriginal source
Analysis to Flag-warning of Automated Hematology Analyzer and Blood-smear Microscopy Examination — Research Paper | ScholarLens