Bio-Rad D10 Automatic Analyzer for Hemoglobin A1c Detection:Preliminary Evaluation
Qiaoping Wu
Abstract
Qiaoping Wu
Abstract
Objective To evaluate the utilization of Bio-Rad D10 automatic analyzer for hemoglobin A1c detection.Methods In accordance with the NCCLS Document EP10-A,a preliminary evaluation of the D10 HbA1c analyzer was made on calibrators at high,middle and low concentrations for a period of 5 days,and the deviation of HbA1c,total imprecision,the slope,carrying pollution,nonlinearity and drift were calculated,and T test was conducted at the same time.Results The high value,medium value and low value deviations were-0.01%,-0.09% and-0.11%,respectively within the permissive ranges.The total imprecision was 1.148%,0.886% and 0.727%,respectively,and they were all less than the permissive 5% ranges of ADA.There were significant differences in the slopes and no significant difference in their intercepts,nonlinearities,carrying pollutions and drifts.Conclusion The precision and accuracy of Bio-Rad D10 automatic analyzer for hemoglobin A1c detection are very good,and its performance indicators consistent with clinical application requirements.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To evaluate the utilization of Bio-Rad D10 automatic analyzer for hemoglobin A1c detection.Methods In accordance with the NCCLS Document EP10-A,a preliminary evaluation of the D10 HbA1c analyzer was made on calibrators at high,middle and low concentrations for a period of 5 days,and the deviation of HbA1c,total imprecision,the slope,carrying pollution,nonlinearity and drift were calculated,and T test was conducted at the same time.Results The high value,medium value and low value deviations were-0.01%,-0.09% and-0.11%,respectively within the permissive ranges.The total imprecision was 1.148%,0.886% and 0.727%,respectively,and they were all less than the permissive 5% ranges of ADA.There were significant differences in the slopes and no significant difference in their intercepts,nonlinearities,carrying pollutions and drifts.Conclusion The precision and accuracy of Bio-Rad D10 automatic analyzer for hemoglobin A1c detection are very good,and its performance indicators consistent with clinical application requirements.
Key concepts: Spectrum analyzer, Medicine, Hemoglobin, Biomedical engineering, Nuclear medicine, Statistics, Internal medicine, Optics