The Accuracy Evaluation of Nucleated Red Blood Cell Count Using the Sysmex XE-2100 Automated Hematology Analyzer
Xueying Wang
Abstract
Xueying Wang
Abstract
Objective To evaluate the accuracy of Sysmex XE-2100 automated hematology analyzer on counting nucleated red blood cell(NRBC).Methods Nucleated red blood cells were counted from 69 positive specimens by Sysmex XE-2100automatedhematologyanalyzerandcomparedwiththemanualmethodwithmicroscopy.Results Therewasa good correlation between the manual and automatic method(r=0.992).No significant difference was observed between the counting results of two methods(P0.05).When the peripheral NRBC appeared,the counting of WBC was significantly higher than the value obtained when the NRBC was depleted and the proportion of lymphocytes was significantly lower(P0.05).While there was no difference among the red blood cells,hemoglobin and platelets(P0.05).Conclusion The NRBC counting by the XE-2100 automated hematology analyzer is sensitive,quick,accurate and reliable.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To evaluate the accuracy of Sysmex XE-2100 automated hematology analyzer on counting nucleated red blood cell(NRBC).Methods Nucleated red blood cells were counted from 69 positive specimens by Sysmex XE-2100automatedhematologyanalyzerandcomparedwiththemanualmethodwithmicroscopy.Results Therewasa good correlation between the manual and automatic method(r=0.992).No significant difference was observed between the counting results of two methods(P0.05).When the peripheral NRBC appeared,the counting of WBC was significantly higher than the value obtained when the NRBC was depleted and the proportion of lymphocytes was significantly lower(P0.05).While there was no difference among the red blood cells,hemoglobin and platelets(P0.05).Conclusion The NRBC counting by the XE-2100 automated hematology analyzer is sensitive,quick,accurate and reliable.
Key concepts: Hematology analyzer, Nucleated Red Blood Cell, Cell counting, Hematology, Peripheral blood, Hemoglobin, Blood cell, Significant difference