The accuracy evaluation of nucleated red blood cell count determined by SYSMEX XN-1000 automatic hematology analyzer

Wang Yefe

Open publisher page 0 citations

Abstract

Objective To investigate the accuracy of nucleated red blood cell( NRBC) count determined by SYSMEX XN-1000 automatic hematology analyzer( XN-1000). Methods A total of 66 peripheral blood samples were collected and determined for NRBC by XN-1000,and the results of NRBC% were compared with those by flow cytometry( FCM) and manual microscopy for the same sample. The manual microscopy was as golden standard. Results There was a good correlation for NRBC between XN-1000 and microscopy( r = 0. 927),but the correlation was lower than that between FCM and manual microscopy( r = 0. 956). The correlation between XN-1000 and FCM was the best( r = 0. 994). The results of XN-1000 and FCM with manual microscopy had no statistical significance( P 0. 05). XN-1000 for NRBC% had a better specificity( 95. 7%) than FCM. Conclusions XN-1000 automatic hematology analyzer has good performance for detecting NRBC in peripheral blood rapidly,accurately and effectively.

About this research paper

What this paper is about

Objective To investigate the accuracy of nucleated red blood cell( NRBC) count determined by SYSMEX XN-1000 automatic hematology analyzer( XN-1000). Methods A total of 66 peripheral blood samples were collected and determined for NRBC by XN-1000,and the results of NRBC% were compared with those by flow cytometry( FCM) and manual microscopy for the same sample. The manual microscopy was as golden standard. Results There was a good correlation for NRBC between XN-1000 and microscopy( r = 0. 927),but the correlation was lower than that between FCM and manual microscopy( r = 0. 956). The correlation between XN-1000 and FCM was the best( r = 0. 994). The results of XN-1000 and FCM with manual microscopy had no statistical significance( P 0. 05). XN-1000 for NRBC% had a better specificity( 95. 7%) than FCM. Conclusions XN-1000 automatic hematology analyzer has good performance for detecting NRBC in peripheral blood rapidly,accurately and effectively.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To investigate the accuracy of nucleated red blood cell( NRBC) count determined by SYSMEX XN-1000 automatic hematology analyzer( XN-1000). Methods A total of 66 peripheral blood samples were collected and determined for NRBC by XN-1000,and the results of NRBC% were compared with those by flow cytometry( FCM) and manual microscopy for the same sample. The manual microscopy was as golden standard. Results There was a good correlation for NRBC between XN-1000 and microscopy( r = 0. 927),but the correlation was lower than that between FCM and manual microscopy( r = 0. 956). The correlation between XN-1000 and FCM was the best( r = 0. 994). The results of XN-1000 and FCM with manual microscopy had no statistical significance( P 0. 05). XN-1000 for NRBC% had a better specificity( 95. 7%) than FCM. Conclusions XN-1000 automatic hematology analyzer has good performance for detecting NRBC in peripheral blood rapidly,accurately and effectively.

Key concepts: Hematology analyzer, Nucleated Red Blood Cell, Peripheral blood, Flow cytometry, Hematology, Nuclear medicine, Medicine, Pathology

Related papers

Back to paper searchBrowse research topicsOriginal source
The accuracy evaluation of nucleated red blood cell count determined by SYSMEX XN-1000 automatic hematology analyzer — Research Paper | ScholarLens