2012Huaxi yixueRequires access

Comparison of Three Different Instruments Measuring Central Corneal Thickness before and after Laser in situ Keratomileusis

KE Hong-qi

Open publisher page 0 citations

Abstract

Objective To compare the difference in measurements of central corneal thickness(CCT) using A-scan,OrbscanⅡand Pentacam before and after laser in situ keratomileusis(LASIK).Methods Between October 2010 and March 2011,the CCT of 137 patients(274 eyes) were measured by A-scan,OrbscanⅡ and Pentacam,and the results were analyzed by paired t-tests and Pearson correlation.Results Before LASIK,the values of CCT measured by A-scan,OrbscanⅡ and Pentacam were(526.6 ± 34.1),(516.6 ± 34.2),and(539.8 ± 31.5) μm respectively;paired t-tests showed the CCT values obtained with Pentacam were significantly higher than those with other methods(P0.05),but there were no statistical significant differences between OrbscanⅡand Pentacam measurements(P0.05).Six months after LASIK,the values of CCT measured by A-scan,OrbscanⅡand Pentacam was(448.2 ± 48.5),(391.9 ± 58.5),and(451.5 ± 46.4) μm respectively;the CCT values obtained with A-scan and Pentacam didn't differ much from each other(P0.05),and the CCT values obtained with OrbscanⅡwere lower than those obtained with A-scan and Pentacam.There was a high correlation between A-scan and Pentacam measurements.Conclusion The these methods measuring CCT could not be used interchangeably,and A-scan and Pentacam after LASIK were more precise than OrbscanⅡ.

About this research paper

What this paper is about

Objective To compare the difference in measurements of central corneal thickness(CCT) using A-scan,OrbscanⅡand Pentacam before and after laser in situ keratomileusis(LASIK).Methods Between October 2010 and March 2011,the CCT of 137 patients(274 eyes) were measured by A-scan,OrbscanⅡ and Pentacam,and the results were analyzed by paired t-tests and Pearson correlation.Results Before LASIK,the values of CCT measured by A-scan,OrbscanⅡ and Pentacam were(526.6 ± 34.1),(516.6 ± 34.2),and(539.8 ± 31.5) μm respectively;paired t-tests showed the CCT values obtained with Pentacam were significantly higher than those with other methods(P0.05),but there were no statistical significant differences between OrbscanⅡand Pentacam measurements(P0.05).Six months after LASIK,the values of CCT measured by A-scan,OrbscanⅡand Pentacam was(448.2 ± 48.5),(391.9 ± 58.5),and(451.5 ± 46.4) μm respectively;the CCT values obtained with A-scan and Pentacam didn't differ much from each other(P0.05),and the CCT values obtained with OrbscanⅡwere lower than those obtained with A-scan and Pentacam.There was a high correlation between A-scan and Pentacam measurements.Conclusion The these methods measuring CCT could not be used interchangeably,and A-scan and Pentacam after LASIK were more precise than OrbscanⅡ.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To compare the difference in measurements of central corneal thickness(CCT) using A-scan,OrbscanⅡand Pentacam before and after laser in situ keratomileusis(LASIK).Methods Between October 2010 and March 2011,the CCT of 137 patients(274 eyes) were measured by A-scan,OrbscanⅡ and Pentacam,and the results were analyzed by paired t-tests and Pearson correlation.Results Before LASIK,the values of CCT measured by A-scan,OrbscanⅡ and Pentacam were(526.6 ± 34.1),(516.6 ± 34.2),and(539.8 ± 31.5) μm respectively;paired t-tests showed the CCT values obtained with Pentacam were significantly higher than those with other methods(P0.05),but there were no statistical significant differences between OrbscanⅡand Pentacam measurements(P0.05).Six months after LASIK,the values of CCT measured by A-scan,OrbscanⅡand Pentacam was(448.2 ± 48.5),(391.9 ± 58.5),and(451.5 ± 46.4) μm respectively;the CCT values obtained with A-scan and Pentacam didn't differ much from each other(P0.05),and the CCT values obtained with OrbscanⅡwere lower than those obtained with A-scan and Pentacam.There was a high correlation between A-scan and Pentacam measurements.Conclusion The these methods measuring CCT could not be used interchangeably,and A-scan and Pentacam after LASIK were more precise than OrbscanⅡ.

Key concepts: Keratomileusis, Medicine, LASIK, Ophthalmology, Significant difference, Cornea, Internal medicine

Related papers

Back to paper searchBrowse research topicsOriginal source
Comparison of Three Different Instruments Measuring Central Corneal Thickness before and after Laser in situ Keratomileusis — Research Paper | ScholarLens