Comparison of Central Corneal Thickness Measurements with Pentacam, Orbscan II, and Ultrasound Pachymeter

Abbasali Yekta, Hassan Hashemi, Khabaz Khoub Mahdi, A Doustdar, Shiva Mehravaran, Javad Heravian, Akbar Fotouhi

Open publisher page 5 citations

Abstract

Purpose: To compare thickness of central cornea measured using Pentacam, Orbscan II, and ultrasound pachymeter Methods: Patients with no history of corneal diseases or systemic diseases affecting eyes, who did not wear contact lens or use eye medications, and who with no previous history of corneal surgery were selected for this study. Central corneal thickness (CCT) was measured by three methods using Pentacam, Orbscan II, and ultrasound pachymeter. Results: Comparison of ultrasound and Orbscan CCT measurements showed a relatively high correlation between these two devices (P<0.001; r=0.891). The 95% limits of agreement (LoA) between these two devices were -42.44 to 20.18 µm. There was also a high correlation between the results obtained through ultrasound and Pentacam (P<0.001; r=0.932). The 95% LoA of CCT with ultrasound and Pentacam were -13.35 to 24.16 µm. There was also a high correlation between CCT measurements carried out by Orbscan and Pentacam (P<0.001) and the 95% LoA were -12.14 to 45.19 µm. Conclusion: The findings of the present study demonstrated high agreements between the CCT readings measured with Orbscan, Pentacam, and ultrasound. The agreement between the Pentacam and ultrasound measurements was higher than that of between Orbscan and ultrasound, making Pentacam a better substitute for ultrasound.

About this research paper

What this paper is about

Purpose: To compare thickness of central cornea measured using Pentacam, Orbscan II, and ultrasound pachymeter Methods: Patients with no history of corneal diseases or systemic diseases affecting eyes, who did not wear contact lens or use eye medications, and who with no previous history of corneal surgery were selected for this study. Central corneal thickness (CCT) was measured by three methods using Pentacam, Orbscan II, and ultrasound pachymeter. Results: Comparison of ultrasound and Orbscan CCT measurements showed a relatively high correlation between these two devices (P<0.001; r=0.891). The 95% limits of agreement (LoA) between these two devices were -42.44 to 20.18 µm. There was also a high correlation between the results obtained through ultrasound and Pentacam (P<0.001; r=0.932). The 95% LoA of CCT with ultrasound and Pentacam were -13.35 to 24.16 µm. There was also a high correlation between CCT measurements carried out by Orbscan and Pentacam (P<0.001) and the 95% LoA were -12.14 to 45.19 µm. Conclusion: The findings of the present study demonstrated high agreements between the CCT readings measured with Orbscan, Pentacam, and ultrasound. The agreement between the Pentacam and ultrasound measurements was higher than that of between Orbscan and ultrasound, making Pentacam a better substitute for ultrasound.

Why it matters

OpenAlex reports 5 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Purpose: To compare thickness of central cornea measured using Pentacam, Orbscan II, and ultrasound pachymeter Methods: Patients with no history of corneal diseases or systemic diseases affecting eyes, who did not wear contact lens or use eye medications, and who with no previous history of corneal surgery were selected for this study. Central corneal thickness (CCT) was measured by three methods using Pentacam, Orbscan II, and ultrasound pachymeter. Results: Comparison of ultrasound and Orbscan CCT measurements showed a relatively high correlation between these two devices (P<0.001; r=0.891). The 95% limits of agreement (LoA) between these two devices were -42.44 to 20.18 µm. There was also a high correlation between the results obtained through ultrasound and Pentacam (P<0.001; r=0.932). The 95% LoA of CCT with ultrasound and Pentacam were -13.35 to 24.16 µm. There was also a high correlation between CCT measurements carried out by Orbscan and Pentacam (P<0.001) and the 95% LoA were -12.14 to 45.19 µm. Conclusion: The findings of the present study demonstrated high agreements between the CCT readings measured with Orbscan, Pentacam, and ultrasound. The agreement between the Pentacam and ultrasound measurements was higher than that of between Orbscan and ultrasound, making Pentacam a better substitute for ultrasound.