Reduction of early elastic recoil by cutting balloon angioplasty as compared to conventional balloon angioplasty.

Katsuhiro Kawaguchi, Taizo Kondo, Takanori Shumiya, Tomohiro Uchikawa, Tomomichi Suzuki, Yoshifumi Awaji, Mitsuhiro Mochizuki

Open publisher page 20 citations

Abstract

BACKGROUND: Restenosis after successful balloon angioplasty remains problematic. Early elastic response after angioplasty is significant when considering the possible development of restenosis. The purpose of this study was to compare early elastic recoil within 10 minutes after successful percutaneous transluminal coronary angioplasty and early lumen loss at 24 hours after angioplasty in a cutting balloon group and a conventional balloon group. METHODS: Extent of early elastic recoil was quantitatively measured as the difference of mean balloon diameter at maximal inflation pressure and minimal luminal diameter after angioplasty in 82 cutting balloon-treated lesions and 51 conventional balloon-treated lesions. RESULTS: Reference diameter and balloon/artery ratio were similar between the cutting balloon and conventional balloon groups (2.89 0.47 mm vs. 2.88 0.60 mm; 1.19 0.11 vs. 1.19 0.13, respectively). Early elastic recoil after angioplasty was significantly smaller in the cutting balloon than the conventional balloon group (0.96 0.40 mm vs. 1.12 0.37 mm, respectively; p = 0.04). Also, the mean amount of lumen loss from 10 minutes after angioplasty to 24 hours after was significantly smaller in the cutting balloon than the conventional balloon group (0.08 0.28 mm vs. 0.20 0.33 mm, respectively; p = 0.02). CONCLUSION: There is significantly less early elastic recoil in the cutting balloon angioplasty than in the conventional balloon angioplasty group. The efficacy of cutting balloon continues 24 hours after angioplasty.

About this research paper

What this paper is about

BACKGROUND: Restenosis after successful balloon angioplasty remains problematic. Early elastic response after angioplasty is significant when considering the possible development of restenosis. The purpose of this study was to compare early elastic recoil within 10 minutes after successful percutaneous transluminal coronary angioplasty and early lumen loss at 24 hours after angioplasty in a cutting balloon group and a conventional balloon group. METHODS: Extent of early elastic recoil was quantitatively measured as the difference of mean balloon diameter at maximal inflation pressure and minimal luminal diameter after angioplasty in 82 cutting balloon-treated lesions and 51 conventional balloon-treated lesions. RESULTS: Reference diameter and balloon/artery ratio were similar between the cutting balloon and conventional balloon groups (2.89 0.47 mm vs. 2.88 0.60 mm; 1.19 0.11 vs. 1.19 0.13, respectively). Early elastic recoil after angioplasty was significantly smaller in the cutting balloon than the conventional balloon group (0.96 0.40 mm vs. 1.12 0.37 mm, respectively; p = 0.04). Also, the mean amount of lumen loss from 10 minutes after angioplasty to 24 hours after was significantly smaller in the cutting balloon than the conventional balloon group (0.08 0.28 mm vs. 0.20 0.33 mm, respectively; p = 0.02). CONCLUSION: There is significantly less early elastic recoil in the cutting balloon angioplasty than in the conventional balloon angioplasty group. The efficacy of cutting balloon continues 24 hours after angioplasty.

Why it matters

OpenAlex reports 20 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

BACKGROUND: Restenosis after successful balloon angioplasty remains problematic. Early elastic response after angioplasty is significant when considering the possible development of restenosis. The purpose of this study was to compare early elastic recoil within 10 minutes after successful percutaneous transluminal coronary angioplasty and early lumen loss at 24 hours after angioplasty in a cutting balloon group and a conventional balloon group. METHODS: Extent of early elastic recoil was quantitatively measured as the difference of mean balloon diameter at maximal inflation pressure and minimal luminal diameter after angioplasty in 82 cutting balloon-treated lesions and 51 conventional balloon-treated lesions. RESULTS: Reference diameter and balloon/artery ratio were similar between the cutting balloon and conventional balloon groups (2.89 0.47 mm vs. 2.88 0.60 mm; 1.19 0.11 vs. 1.19 0.13, respectively). Early elastic recoil after angioplasty was significantly smaller in the cutting balloon than the conventional balloon group (0.96 0.40 mm vs. 1.12 0.37 mm, respectively; p = 0.04). Also, the mean amount of lumen loss from 10 minutes after angioplasty to 24 hours after was significantly smaller in the cutting balloon than the conventional balloon group (0.08 0.28 mm vs. 0.20 0.33 mm, respectively; p = 0.02). CONCLUSION: There is significantly less early elastic recoil in the cutting balloon angioplasty than in the conventional balloon angioplasty group. The efficacy of cutting balloon continues 24 hours after angioplasty.

Key concepts: Balloon, Angioplasty, Elastic recoil, Medicine, Restenosis, Cutting balloon, Lumen (anatomy), Recoil

Related papers

Back to paper searchBrowse research topicsOriginal source
Reduction of early elastic recoil by cutting balloon angioplasty as compared to conventional balloon angioplasty. — Research Paper | ScholarLens