Performance of Abbott Alinity hq hematology analyzer for automated cell counting of body fluids

Pauline Herroelen, Simke Demeester, Serge Damiaens, Anton Evenepoel, Kristin Jochmans

Open publisher page 0 citations

Abstract

INTRODUCTION: Body fluid cell counting and differentiation provide essential information for diagnosis and monitoring of diverse pathologies. We evaluated the performance of the newly launched Abbott Alinity hq hematology analyzer for automated cell counting in body fluids and compared red blood cell (RBC) and total nucleated cell (TNC) counts with the Cell-Dyn Sapphire automated hematology analyzer. Differential counts were compared with microscopic differentiation on cytocentrifuged preparations. METHODS: Background concentration limits, limit of detection (LOD), linearity, imprecision, functional sensitivity and carryover were evaluated. For method comparison, we collected 172 body fluids (17 continuous ambulatory peritoneal dialysis fluids, 56 cerebrospinal fluids and 99 serous fluids). RESULTS: Background concentration limits were ≤1000 cells/μL for RBC counts and ≤3 cells/μL for TNC counts. The LOD was 1000 RBC/µL and 5 TNC/µL. Results from linear regression analysis revealed excellent linearity. Functional sensitivity was 3000 cells/µL for RBC counts and 50 cells/µL for TNC counts. Carryover was 0.6% and 0.1% for TNC and RBC, respectively. The Alinity hq shows good clinical performance. CONCLUSION: We demonstrated comparable performance for body fluid cell counting between the Alinity hq analyzer and the Cell-Dyn Sapphire. The Alinity hq can be very useful as a screening tool for body fluid cell counting.

About this research paper

What this paper is about

INTRODUCTION: Body fluid cell counting and differentiation provide essential information for diagnosis and monitoring of diverse pathologies. We evaluated the performance of the newly launched Abbott Alinity hq hematology analyzer for automated cell counting in body fluids and compared red blood cell (RBC) and total nucleated cell (TNC) counts with the Cell-Dyn Sapphire automated hematology analyzer. Differential counts were compared with microscopic differentiation on cytocentrifuged preparations. METHODS: Background concentration limits, limit of detection (LOD), linearity, imprecision, functional sensitivity and carryover were evaluated. For method comparison, we collected 172 body fluids (17 continuous ambulatory peritoneal dialysis fluids, 56 cerebrospinal fluids and 99 serous fluids). RESULTS: Background concentration limits were ≤1000 cells/μL for RBC counts and ≤3 cells/μL for TNC counts. The LOD was 1000 RBC/µL and 5 TNC/µL. Results from linear regression analysis revealed excellent linearity. Functional sensitivity was 3000 cells/µL for RBC counts and 50 cells/µL for TNC counts. Carryover was 0.6% and 0.1% for TNC and RBC, respectively. The Alinity hq shows good clinical performance. CONCLUSION: We demonstrated comparable performance for body fluid cell counting between the Alinity hq analyzer and the Cell-Dyn Sapphire. The Alinity hq can be very useful as a screening tool for body fluid cell counting.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

INTRODUCTION: Body fluid cell counting and differentiation provide essential information for diagnosis and monitoring of diverse pathologies. We evaluated the performance of the newly launched Abbott Alinity hq hematology analyzer for automated cell counting in body fluids and compared red blood cell (RBC) and total nucleated cell (TNC) counts with the Cell-Dyn Sapphire automated hematology analyzer. Differential counts were compared with microscopic differentiation on cytocentrifuged preparations. METHODS: Background concentration limits, limit of detection (LOD), linearity, imprecision, functional sensitivity and carryover were evaluated. For method comparison, we collected 172 body fluids (17 continuous ambulatory peritoneal dialysis fluids, 56 cerebrospinal fluids and 99 serous fluids). RESULTS: Background concentration limits were ≤1000 cells/μL for RBC counts and ≤3 cells/μL for TNC counts. The LOD was 1000 RBC/µL and 5 TNC/µL. Results from linear regression analysis revealed excellent linearity. Functional sensitivity was 3000 cells/µL for RBC counts and 50 cells/µL for TNC counts. Carryover was 0.6% and 0.1% for TNC and RBC, respectively. The Alinity hq shows good clinical performance. CONCLUSION: We demonstrated comparable performance for body fluid cell counting between the Alinity hq analyzer and the Cell-Dyn Sapphire. The Alinity hq can be very useful as a screening tool for body fluid cell counting.

Key concepts: Hematology analyzer, Cell counting, Body fluid, Detection limit, Cell, Blood cell, Medicine, Pathology