Standardization of a Radiofrequency Ablation Tool in an Ex-Vivo Porcine Liver Model

Gemma Rossi, Maria Chiara Petrone, Gabriele Capurso, Luca Albarello, Sabrina Gloria Giulia Testoni, Lívia Archibugi, Marco Schiavo Lena, Claudio Doglioni, Paolo Giorgio Arcidiacono

Open full text 11 citations

Abstract

(1) Background: Preclinical and clinical data about a novel radiofrequency ablation (RFA) system (STARmed Co, Ltd.; Koyang, Korea) designed to be used under endoscopic ultrasound (EUS) control for pancreatic lesion ablation, are limited, obtained with non-standardized procedures and heterogeneous results. The aim of this study is to standardize the RFA procedure of this system in order to define the optimal ablation power and time. (2) Methods: RFA was performed on an ex-vivo porcine liver at different powers (40, 30, 20, 10 Watts (W)) and times (1, 3, 5, 7, 15 min) with a 1-centimeter monopolar electrode (perfused by chilled solution) positioned on the distal tip of a 19-Gauge needle. A blinded expert pathologist histologically analyzed each ablation area. (3) Results: The size of the total macroscopic ablated area was negatively correlated with ablation power (R −0.74): the largest was obtained at 10 W (p = 4.7 × 10−4) for longer times (R 0.92; p = 8.9 × 10−8). Central histologic coagulative necrosis did not differ among ablation settings (mean size 3.25 mm). External “parenchymal hypochromia” or “diaphanization” resulted the widest at 10 W, for longer times (R 0.8, p = 3.6 × 10−4). (4) Conclusions: The RFA system can produce small sizes of coagulative necrosis, regardless of the setting. Larger areas of diaphanization surrounding the necrosis can be produced at lower powers for longer times.

Open-access reader

About this research paper

What this paper is about

(1) Background: Preclinical and clinical data about a novel radiofrequency ablation (RFA) system (STARmed Co, Ltd.; Koyang, Korea) designed to be used under endoscopic ultrasound (EUS) control for pancreatic lesion ablation, are limited, obtained with non-standardized procedures and heterogeneous results. The aim of this study is to standardize the RFA procedure of this system in order to define the optimal ablation power and time. (2) Methods: RFA was performed on an ex-vivo porcine liver at different powers (40, 30, 20, 10 Watts (W)) and times (1, 3, 5, 7, 15 min) with a 1-centimeter monopolar electrode (perfused by chilled solution) positioned on the distal tip of a 19-Gauge needle. A blinded expert pathologist histologically analyzed each ablation area. (3) Results: The size of the total macroscopic ablated area was negatively correlated with ablation power (R −0.74): the largest was obtained at 10 W (p = 4.7 × 10−4) for longer times (R 0.92; p = 8.9 × 10−8). Central histologic coagulative necrosis did not differ among ablation settings (mean size 3.25 mm). External “parenchymal hypochromia” or “diaphanization” resulted the widest at 10 W, for longer times (R 0.8, p = 3.6 × 10−4). (4) Conclusions: The RFA system can produce small sizes of coagulative necrosis, regardless of the setting. Larger areas of diaphanization surrounding the necrosis can be produced at lower powers for longer times.

Why it matters

OpenAlex reports 11 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

(1) Background: Preclinical and clinical data about a novel radiofrequency ablation (RFA) system (STARmed Co, Ltd.; Koyang, Korea) designed to be used under endoscopic ultrasound (EUS) control for pancreatic lesion ablation, are limited, obtained with non-standardized procedures and heterogeneous results. The aim of this study is to standardize the RFA procedure of this system in order to define the optimal ablation power and time. (2) Methods: RFA was performed on an ex-vivo porcine liver at different powers (40, 30, 20, 10 Watts (W)) and times (1, 3, 5, 7, 15 min) with a 1-centimeter monopolar electrode (perfused by chilled solution) positioned on the distal tip of a 19-Gauge needle. A blinded expert pathologist histologically analyzed each ablation area. (3) Results: The size of the total macroscopic ablated area was negatively correlated with ablation power (R −0.74): the largest was obtained at 10 W (p = 4.7 × 10−4) for longer times (R 0.92; p = 8.9 × 10−8). Central histologic coagulative necrosis did not differ among ablation settings (mean size 3.25 mm). External “parenchymal hypochromia” or “diaphanization” resulted the widest at 10 W, for longer times (R 0.8, p = 3.6 × 10−4). (4) Conclusions: The RFA system can produce small sizes of coagulative necrosis, regardless of the setting. Larger areas of diaphanization surrounding the necrosis can be produced at lower powers for longer times.

Key concepts: Coagulative necrosis, Ablation, Radiofrequency ablation, Medicine, Ex vivo, Ablation zone, Nuclear medicine, Necrosis