Performance verification and evaluation of 4 domestic chemiluminescence systems on 8 tumor markers
Yan Shi, Yancai Wei, Weiling Zheng, Wei Jia-ling, Miaoli Song, Gengchao Zhu, Xun Lu
Abstract
Yan Shi, Yancai Wei, Weiling Zheng, Wei Jia-ling, Miaoli Song, Gengchao Zhu, Xun Lu
Abstract
Objective To validate the performance of 4 domestic chemiluminescence immunoassay (CLIA) systems on 8 tumor markers quantitative assay kits. Methods Four domestic CLIA systems were randomly marked as A, B, C, D and 8 tumor markers, including carbohydrate antigen (CA)125, CA15-3, CA19-9, ferritin (Fer), alpha-fetoprotein (AFP), carcinoembryonic antigen (CEA), prostate-specific antigen (PSA) and free PSA (fPSA) were determined. According to the standard of Clinical and Laboratory Standards Institute (CLSI), the precision, methodological comparison and analytical measure range of 4 systems were validated. Clinical serum samples were obtained from patients in Suzhou Hospital. According to the CLSI EP9-A3 protocol, imported equipment was used as the reference system. The biases of medical decision points were assumed, and Pearson correlation analysis and Spearman correlation analysis were used to analyze the data. Results The precision verification of CA125 and PSA on A, CA125 and AFP on B, CA125, CEA, AFP and PSA on C, and all 8 tumor markers on D could meet the laboratory quality control requirements. The correlations of the test results between A-D and the imported equipment were significant (all P<0.05) with the correlation coefficients 0.79-0.99, 0.47-0.99, 0.90-0.98 and 0.78-1.00, respectively, and the number of acceptable tests at the level of medical decision was 5, 2, 5, 4. All tests were certified to meet the analytical measure range validation. Conclusions The detection performance of 4 domestic CLIA systems for all 8 tumor markers are different. The performance of domestic CLIA systems should be tested when choosing one that can meet laboratory quality control requirements. Key words: Chemiluminescent measurements; Immunoassay; Tumor markers, biological
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To validate the performance of 4 domestic chemiluminescence immunoassay (CLIA) systems on 8 tumor markers quantitative assay kits. Methods Four domestic CLIA systems were randomly marked as A, B, C, D and 8 tumor markers, including carbohydrate antigen (CA)125, CA15-3, CA19-9, ferritin (Fer), alpha-fetoprotein (AFP), carcinoembryonic antigen (CEA), prostate-specific antigen (PSA) and free PSA (fPSA) were determined. According to the standard of Clinical and Laboratory Standards Institute (CLSI), the precision, methodological comparison and analytical measure range of 4 systems were validated. Clinical serum samples were obtained from patients in Suzhou Hospital. According to the CLSI EP9-A3 protocol, imported equipment was used as the reference system. The biases of medical decision points were assumed, and Pearson correlation analysis and Spearman correlation analysis were used to analyze the data. Results The precision verification of CA125 and PSA on A, CA125 and AFP on B, CA125, CEA, AFP and PSA on C, and all 8 tumor markers on D could meet the laboratory quality control requirements. The correlations of the test results between A-D and the imported equipment were significant (all P<0.05) with the correlation coefficients 0.79-0.99, 0.47-0.99, 0.90-0.98 and 0.78-1.00, respectively, and the number of acceptable tests at the level of medical decision was 5, 2, 5, 4. All tests were certified to meet the analytical measure range validation. Conclusions The detection performance of 4 domestic CLIA systems for all 8 tumor markers are different. The performance of domestic CLIA systems should be tested when choosing one that can meet laboratory quality control requirements. Key words: Chemiluminescent measurements; Immunoassay; Tumor markers, biological
Key concepts: Carcinoembryonic antigen, External quality assessment, Immunoassay, Medicine, Prostate-specific antigen, Chemiluminescent immunoassay, Spearman's rank correlation coefficient, Oncology