2013The Journal of Medical Theory and PracticeRequires access

The Comparative Analysis of Conventional Tumor Markers in the Two Detection Systems between Roche E411 and E601

Ni Tang

Open publisher page 0 citations

Abstract

Objective:To investigate the comparability of test results of two electrochemiluminescence immunoassay analyzer between Roche Cobas E411and Cobas E601.Methods:According to United States Clinical and Laboratory Standards Institute(CLSI)EP9-A2document,E411was acted as comparison method(X)and E601was as the experimental method(Y).Alpha fetoprotein(AFP),carcinoembryonic antigen(CEA),carbohydrate antigen 125(CA125)and carbohydrate antigen 199(CA199)levels of patient samples were tested and the results were analyzed.Results:The correlation between the two systems is good and the bias is acceptable in the medical decision level.Conclusion:The comparative analysis and bias evaluation systems should be carried out to ensure the comparability of test results when the same items are detected with different detection systems.

About this research paper

What this paper is about

Objective:To investigate the comparability of test results of two electrochemiluminescence immunoassay analyzer between Roche Cobas E411and Cobas E601.Methods:According to United States Clinical and Laboratory Standards Institute(CLSI)EP9-A2document,E411was acted as comparison method(X)and E601was as the experimental method(Y).Alpha fetoprotein(AFP),carcinoembryonic antigen(CEA),carbohydrate antigen 125(CA125)and carbohydrate antigen 199(CA199)levels of patient samples were tested and the results were analyzed.Results:The correlation between the two systems is good and the bias is acceptable in the medical decision level.Conclusion:The comparative analysis and bias evaluation systems should be carried out to ensure the comparability of test results when the same items are detected with different detection systems.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective:To investigate the comparability of test results of two electrochemiluminescence immunoassay analyzer between Roche Cobas E411and Cobas E601.Methods:According to United States Clinical and Laboratory Standards Institute(CLSI)EP9-A2document,E411was acted as comparison method(X)and E601was as the experimental method(Y).Alpha fetoprotein(AFP),carcinoembryonic antigen(CEA),carbohydrate antigen 125(CA125)and carbohydrate antigen 199(CA199)levels of patient samples were tested and the results were analyzed.Results:The correlation between the two systems is good and the bias is acceptable in the medical decision level.Conclusion:The comparative analysis and bias evaluation systems should be carried out to ensure the comparability of test results when the same items are detected with different detection systems.

Key concepts: Medicine, Carcinoembryonic antigen, Comparability, Immunoassay, Spectrum analyzer, Chromatography, Oncology, Internal medicine