A biomechanical study of patellar tension plating system for patellar fractures
Tianyi Zhou, Shixin Guo, Baitong Li, Xinguang Han
Abstract
Tianyi Zhou, Shixin Guo, Baitong Li, Xinguang Han
Abstract
Objective To investigate biomechanical properties of the patella tension plating system in order to provide a theoretical basis for its clinical application. Methods Thirty-six models of artificial patella were randomly divided into 3 groups (n=12). After transverse patellar fractures were created in the models, the 3 groups were subjected to fixation respectively with Kirschner wire tension band (tension band group), patellar concentrator (concentrator group) and patellar tension plate (tension plate group). Next, 6 specimens from each group were placed on a mechanical testing machine to measure the fracture displacements after 100 cycles of simulated knee flexion and extension movements. Tensile strength tests were performed on the remaining 6 specimens in each group to measure the maximum load at fixation failure. Results The fracture displacement in the tension plate group (0.40±0.26 mm) was significantly smaller than those in the tension band group (2.58±0.72 mm) and in the concentrator group (1.25±0.74 mm) (P<0.05); the maximum load at fixation failure in the tension plate group (1,709±206 N) was significantly greater than those in the tension band group (581±122 N) and in the concentrator group (1,003±211 N) (P< 0.05). Conclusion As a new treatment for patellar fractures, the patellar tension plating system can perform better in biomechanical properties than Kirschner wire tension band and patellar concentrator. Key words: Patella; Fractures, bone; Fracture fixation, internal; Internal fixator; Biomechanics
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To investigate biomechanical properties of the patella tension plating system in order to provide a theoretical basis for its clinical application. Methods Thirty-six models of artificial patella were randomly divided into 3 groups (n=12). After transverse patellar fractures were created in the models, the 3 groups were subjected to fixation respectively with Kirschner wire tension band (tension band group), patellar concentrator (concentrator group) and patellar tension plate (tension plate group). Next, 6 specimens from each group were placed on a mechanical testing machine to measure the fracture displacements after 100 cycles of simulated knee flexion and extension movements. Tensile strength tests were performed on the remaining 6 specimens in each group to measure the maximum load at fixation failure. Results The fracture displacement in the tension plate group (0.40±0.26 mm) was significantly smaller than those in the tension band group (2.58±0.72 mm) and in the concentrator group (1.25±0.74 mm) (P<0.05); the maximum load at fixation failure in the tension plate group (1,709±206 N) was significantly greater than those in the tension band group (581±122 N) and in the concentrator group (1,003±211 N) (P< 0.05). Conclusion As a new treatment for patellar fractures, the patellar tension plating system can perform better in biomechanical properties than Kirschner wire tension band and patellar concentrator. Key words: Patella; Fractures, bone; Fracture fixation, internal; Internal fixator; Biomechanics
Key concepts: Patella fracture, Medicine, Fixation (population genetics), Orthodontics, Biomechanics, Tension band wiring, Internal fixation, Tension (geology)