The Influence of Corneal Biomechanical Properties on Intraocular Pressure Measurements Using a Rebound Self-tonometer
Lyndsay Brown, William Foulsham, Savva Pronin, Andrew J. Tatham
Abstract
Lyndsay Brown, William Foulsham, Savva Pronin, Andrew J. Tatham
Abstract
PURPOSE: The purpose of this study was to examine the effect of corneal biomechanical properties on intraocular pressure (IOP) measurements obtained using a rebound self-tonometer (Icare HOME) compared with Goldmann applanation tonometry (GAT). METHODS: An observational study of 100 patients with glaucoma or ocular hypertension. All had a comprehensive ophthalmic examination and standard automated perimetry. IOP was assessed by GAT, Icare HOME and Ocular Response Analyzer, which was also used to assess corneal hysteresis (CH) and corneal resistance factor (CRF). Central corneal thickness (CCT) was recorded. RESULTS: Mean (±SD) IOP measurements were 14.3±3.9 and 11.7±4.7 mm Hg using GAT and Icare HOME, respectively. Average CCT, CRF, and CH were 534.5±37.3 μm, 9.0±1.7 mm Hg, and 9.4±1.5 mm Hg, respectively. The mean difference between Icare HOME and GAT was -2.66±3.13 mm Hg, with 95% limits of agreement of -8.80 to 3.48 mm Hg, however, there was evidence of proportional bias. There was negative correlation between IOP and CH [5.17 mm Hg higher Icare HOME IOP (P=0.041, R=0.029) and 7.23 mm Hg higher GAT IOP (P=0.008, R=0.080) for each 10 mm Hg lower CH], whereas thinner CCT was significantly associated with lower IOP (P<0.001, R=0.14 for Icare HOME and P<0.001, R=0.08 for GAT). In multivariable analysis, although CRF and CH remained associated with IOP measured using either GAT or Icare HOME, CCT was no longer significant. CONCLUSION: IOP measurements obtained using a self-tonometer, similar to GAT, were more influenced by overall corneal biomechanics than CCT.
OpenAlex reports 43 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
PURPOSE: The purpose of this study was to examine the effect of corneal biomechanical properties on intraocular pressure (IOP) measurements obtained using a rebound self-tonometer (Icare HOME) compared with Goldmann applanation tonometry (GAT). METHODS: An observational study of 100 patients with glaucoma or ocular hypertension. All had a comprehensive ophthalmic examination and standard automated perimetry. IOP was assessed by GAT, Icare HOME and Ocular Response Analyzer, which was also used to assess corneal hysteresis (CH) and corneal resistance factor (CRF). Central corneal thickness (CCT) was recorded. RESULTS: Mean (±SD) IOP measurements were 14.3±3.9 and 11.7±4.7 mm Hg using GAT and Icare HOME, respectively. Average CCT, CRF, and CH were 534.5±37.3 μm, 9.0±1.7 mm Hg, and 9.4±1.5 mm Hg, respectively. The mean difference between Icare HOME and GAT was -2.66±3.13 mm Hg, with 95% limits of agreement of -8.80 to 3.48 mm Hg, however, there was evidence of proportional bias. There was negative correlation between IOP and CH [5.17 mm Hg higher Icare HOME IOP (P=0.041, R=0.029) and 7.23 mm Hg higher GAT IOP (P=0.008, R=0.080) for each 10 mm Hg lower CH], whereas thinner CCT was significantly associated with lower IOP (P<0.001, R=0.14 for Icare HOME and P<0.001, R=0.08 for GAT). In multivariable analysis, although CRF and CH remained associated with IOP measured using either GAT or Icare HOME, CCT was no longer significant. CONCLUSION: IOP measurements obtained using a self-tonometer, similar to GAT, were more influenced by overall corneal biomechanics than CCT.
Key concepts: Medicine, Intraocular pressure, Goldmann Applanation Tonometer, Ophthalmology, Glaucoma, Ocular hypertension, Applanation tonometry, Blood pressure