A Comparison of Intraocular Pressure Measurements with the XPERT Noncontact Tonometer and Goldmann Applanation Tonometry
Karen S. Popovich, M. Bruce Shields
Abstract
Karen S. Popovich, M. Bruce Shields
Abstract
PURPOSE: The purpose of this study was to compare intraocular pressure measurements obtained with XPERT, an upgraded model of the noncontact tonometer, with those from Goldmann applanation tonometry. METHODS: An average of three intraocular pressure measurements, generated by the XPERT, was compared with a single reading from a Goldmann applanation tonometer in 421 eyes of 220 patients. RESULTS: The standard deviation of the matched-pair differences (Sdiff) for the entire population was 2.3 mm Hg, whereas in a subset of 95 eyes with intraocular pressures of > or = 20 mm Hg by Goldmann applanation tonometry, the Sdiff was 3.3 mm Hg. CONCLUSIONS: The XPERT was found to provide reliable intraocular pressure measurements, especially within the normal range.
OpenAlex reports 27 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
PURPOSE: The purpose of this study was to compare intraocular pressure measurements obtained with XPERT, an upgraded model of the noncontact tonometer, with those from Goldmann applanation tonometry. METHODS: An average of three intraocular pressure measurements, generated by the XPERT, was compared with a single reading from a Goldmann applanation tonometer in 421 eyes of 220 patients. RESULTS: The standard deviation of the matched-pair differences (Sdiff) for the entire population was 2.3 mm Hg, whereas in a subset of 95 eyes with intraocular pressures of > or = 20 mm Hg by Goldmann applanation tonometry, the Sdiff was 3.3 mm Hg. CONCLUSIONS: The XPERT was found to provide reliable intraocular pressure measurements, especially within the normal range.
Key concepts: Intraocular pressure, Medicine, Goldmann Applanation Tonometer, Applanation tonometry, Ophthalmology, Glaucoma, Optometry, Blood pressure