Comparison of Impedance Platelet Count by Sysmex XE-5000

Chaicharoen Tantanate, Noppadol Siritanaratkul

Open full text 0 citations

Abstract

Background: Red blood cell (RBC) abnormalities are frequently found among thalassemia patients and have been reported to be a cause of spuriously high platelet counts. Objective: To compare impedance platelet count by Sysmex XE-5000 (PLT-XE) and Beckman Coulter LH 780 (PLT-LH) with optical fluorescent platelet count (PLT-O) in thalassemia patients. Methods: EDTA blood specimens from 152 thalassemia patients were analyzed by routine mode of PLT-XE and PLT-LH. The PLT-O run on reticulocyte mode of Sysmex XE-5000 was used as a reference method. Spearman’s rank correlation coefficient (rs ) and Bland-Altman plot were analyzed. Results: Correlations between impedance and PLT-O methods were excellent, as demonstrated by rs of 0.95 and 0.97 for PLT-XE and PLT-LH, respectively. However, there was no agreement between impedance and PLT-O methods. There were differences in biases of PLT-XE between thalassemia minor and both thalassemia intermedia and major. Compared to PLT-O, the concordance rates of diagnosis for platelet count abnormality were 94.7% and 96.7% for PLT-XE and PLT-LH, respectively. Conclusion: Compared to PLT-O, impedance method was associated with spuriously high platelet counts in thalassemia patients. Careful blood smear examination and/or accurate assay should be used in cases suspected of spuriously high platelet count.

About this research paper

What this paper is about

Background: Red blood cell (RBC) abnormalities are frequently found among thalassemia patients and have been reported to be a cause of spuriously high platelet counts. Objective: To compare impedance platelet count by Sysmex XE-5000 (PLT-XE) and Beckman Coulter LH 780 (PLT-LH) with optical fluorescent platelet count (PLT-O) in thalassemia patients. Methods: EDTA blood specimens from 152 thalassemia patients were analyzed by routine mode of PLT-XE and PLT-LH. The PLT-O run on reticulocyte mode of Sysmex XE-5000 was used as a reference method. Spearman’s rank correlation coefficient (rs ) and Bland-Altman plot were analyzed. Results: Correlations between impedance and PLT-O methods were excellent, as demonstrated by rs of 0.95 and 0.97 for PLT-XE and PLT-LH, respectively. However, there was no agreement between impedance and PLT-O methods. There were differences in biases of PLT-XE between thalassemia minor and both thalassemia intermedia and major. Compared to PLT-O, the concordance rates of diagnosis for platelet count abnormality were 94.7% and 96.7% for PLT-XE and PLT-LH, respectively. Conclusion: Compared to PLT-O, impedance method was associated with spuriously high platelet counts in thalassemia patients. Careful blood smear examination and/or accurate assay should be used in cases suspected of spuriously high platelet count.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Background: Red blood cell (RBC) abnormalities are frequently found among thalassemia patients and have been reported to be a cause of spuriously high platelet counts. Objective: To compare impedance platelet count by Sysmex XE-5000 (PLT-XE) and Beckman Coulter LH 780 (PLT-LH) with optical fluorescent platelet count (PLT-O) in thalassemia patients. Methods: EDTA blood specimens from 152 thalassemia patients were analyzed by routine mode of PLT-XE and PLT-LH. The PLT-O run on reticulocyte mode of Sysmex XE-5000 was used as a reference method. Spearman’s rank correlation coefficient (rs ) and Bland-Altman plot were analyzed. Results: Correlations between impedance and PLT-O methods were excellent, as demonstrated by rs of 0.95 and 0.97 for PLT-XE and PLT-LH, respectively. However, there was no agreement between impedance and PLT-O methods. There were differences in biases of PLT-XE between thalassemia minor and both thalassemia intermedia and major. Compared to PLT-O, the concordance rates of diagnosis for platelet count abnormality were 94.7% and 96.7% for PLT-XE and PLT-LH, respectively. Conclusion: Compared to PLT-O, impedance method was associated with spuriously high platelet counts in thalassemia patients. Careful blood smear examination and/or accurate assay should be used in cases suspected of spuriously high platelet count.

Key concepts: Platelet, Thalassemia, Medicine, Gastroenterology, Nuclear medicine, Reticulocyte, Complete blood count, Internal medicine

Related papers

Back to paper searchBrowse research topicsOriginal source
Comparison of Impedance Platelet Count by Sysmex XE-5000 — Research Paper | ScholarLens