Feline Differential Leukocyte Count with ProCyte Dx: Frequency and Severity of a Neutrophil-Lymphocyte Error and How to Avoid It
Harold Tvedten, V. Andersson, Inger Lilliehöök
Abstract
Open-access reader
Harold Tvedten, V. Andersson, Inger Lilliehöök
Abstract
Open-access reader
BACKGROUND: Erroneous neutrophil and lymphocyte counts from analysis of feline blood samples were transferred directly into the hospital information system from the ProCyte Dx hematology instrument in our after-hours laboratory. Errors usually were not detected by the users. HYPOTHESIS/OBJECTIVES: To quantify the frequency and severity of errors associated with the ProCyte Dx analyzer and to identify methods to avoid the errors. ANIMALS: One-hundred six EDTA blood samples routinely submitted from feline hospital patients were analyzed. METHODS: ProCyte differential leukocyte counts were compared to 2 reference methods: Advia 2120 hematology instrument and manual enumeration. Limits for unacceptable deviation from the reference methods were defined as 18 for % lymphocytes and 23 for % neutrophils. RESULTS: Fourteen of 106 samples had unacceptable errors for both lymphocytes and neutrophils compared to both reference methods. Median % lymphocytes in those 14 samples were 11.2, 15.0, and 53.0% for Advia, manual, and ProCyte, respectively. Median % neutrophils were 85.4, 81.5, and 34.2% for Advia, manual, and ProCyte, respectively. All errors were avoided by rejecting automated ProCyte differential leukocyte results whenever the dot plot appeared clearly incorrect, but only 9 of these 14 samples had a ProCyte WBC distribution error flag. CONCLUSIONS AND CLINICAL IMPORTANCE: Results reported by ProCyte had markedly falsely increased lymphocyte and decreased neutrophil counts in 13% of feline patient samples. Users must reject automated differential leukocyte count results when the WBC dot plot appears overtly incorrect. Rejection based only on ProCyte WBC error flag was insufficient.
OpenAlex reports 17 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
BACKGROUND: Erroneous neutrophil and lymphocyte counts from analysis of feline blood samples were transferred directly into the hospital information system from the ProCyte Dx hematology instrument in our after-hours laboratory. Errors usually were not detected by the users. HYPOTHESIS/OBJECTIVES: To quantify the frequency and severity of errors associated with the ProCyte Dx analyzer and to identify methods to avoid the errors. ANIMALS: One-hundred six EDTA blood samples routinely submitted from feline hospital patients were analyzed. METHODS: ProCyte differential leukocyte counts were compared to 2 reference methods: Advia 2120 hematology instrument and manual enumeration. Limits for unacceptable deviation from the reference methods were defined as 18 for % lymphocytes and 23 for % neutrophils. RESULTS: Fourteen of 106 samples had unacceptable errors for both lymphocytes and neutrophils compared to both reference methods. Median % lymphocytes in those 14 samples were 11.2, 15.0, and 53.0% for Advia, manual, and ProCyte, respectively. Median % neutrophils were 85.4, 81.5, and 34.2% for Advia, manual, and ProCyte, respectively. All errors were avoided by rejecting automated ProCyte differential leukocyte results whenever the dot plot appeared clearly incorrect, but only 9 of these 14 samples had a ProCyte WBC distribution error flag. CONCLUSIONS AND CLINICAL IMPORTANCE: Results reported by ProCyte had markedly falsely increased lymphocyte and decreased neutrophil counts in 13% of feline patient samples. Users must reject automated differential leukocyte count results when the WBC dot plot appears overtly incorrect. Rejection based only on ProCyte WBC error flag was insufficient.
Key concepts: Hematology analyzer, Medicine, Leukocyte Counts, Lymphocyte, Complete blood count, Blood count, Absolute neutrophil count, Reference values