CYCLIC TESTING OF MODIFIED BUNNELL PULLOUT SUTURES AND MICRO-MITEK SUTURE ANCHORS IN FDP TENDON DISTAL FIXATION
Kim Latendresse, Eddy Dona, P.J. Scougal, Mark C. Gillies, William Robert Walsh
Abstract
Kim Latendresse, Eddy Dona, P.J. Scougal, Mark C. Gillies, William Robert Walsh
Abstract
Lacerations of the FDP tendon in zone one may be reattached to bone with a modified Bunnell pullout suture or with suture anchors. Eleven cadaveric fingers were submitted to cyclical testing of five hundred cycles with either a modified Bunnell pullout suture of 3-0 polypropylene or a micro-Mitek suture anchor with 3-0 Ethibond. Gap formation was 6.6mm in the modified Bunnell group and 2.0mm in the micro-Mitek group (p Distal zone one FDP tendon lacerations are usually re-attached to bone by a modified Bunnell pullout suture of 3-0 polypropylene. This treatment may lead to moderate to severe losses of DIP joint motion in up to 50% of patients. Suture anchors have recently been introduced as a fixation alternative. Cyclical testing simulating five days of a passive mobilisation protocol was used to compare the Micro-Mitek anchor to the modified-Bunnell pullout suture in FDP tendon fixation. Eleven cadaveric fingers FDP tendons were repaired to bone using a modified Bunnell pullout suture of 3-0 polypropylene or a micro-Mitek anchor with 3-0 Ethibond. Testing was done from 2N to 15N at 5N/sec, for a total of five hundred cycles. Gap formation at the tendon bone interface was measured. Load-to-failure was performed on all specimens. No specimens failed during cyclic testing. Gap formation was 6.6mm (SD 1.2, range 4.9–8.2mm) and 2.0mm (SD = 0.4, range 1.7–2.7mm) for the pullout technique and the micro-Mitek anchor repair respectively (p This data suggests that both fixation techniques may be adequate to sustain five days of simulated passive rehabilitation therapy. Significant gap formation in the modified Bunnell pullout group is of concern although this needs to be correlated in the clinical setting. The lower failure rate of the micro-Mitek group may leave a narrow margin of safety for passive rehabilitation.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Lacerations of the FDP tendon in zone one may be reattached to bone with a modified Bunnell pullout suture or with suture anchors. Eleven cadaveric fingers were submitted to cyclical testing of five hundred cycles with either a modified Bunnell pullout suture of 3-0 polypropylene or a micro-Mitek suture anchor with 3-0 Ethibond. Gap formation was 6.6mm in the modified Bunnell group and 2.0mm in the micro-Mitek group (p Distal zone one FDP tendon lacerations are usually re-attached to bone by a modified Bunnell pullout suture of 3-0 polypropylene. This treatment may lead to moderate to severe losses of DIP joint motion in up to 50% of patients. Suture anchors have recently been introduced as a fixation alternative. Cyclical testing simulating five days of a passive mobilisation protocol was used to compare the Micro-Mitek anchor to the modified-Bunnell pullout suture in FDP tendon fixation. Eleven cadaveric fingers FDP tendons were repaired to bone using a modified Bunnell pullout suture of 3-0 polypropylene or a micro-Mitek anchor with 3-0 Ethibond. Testing was done from 2N to 15N at 5N/sec, for a total of five hundred cycles. Gap formation at the tendon bone interface was measured. Load-to-failure was performed on all specimens. No specimens failed during cyclic testing. Gap formation was 6.6mm (SD 1.2, range 4.9–8.2mm) and 2.0mm (SD = 0.4, range 1.7–2.7mm) for the pullout technique and the micro-Mitek anchor repair respectively (p This data suggests that both fixation techniques may be adequate to sustain five days of simulated passive rehabilitation therapy. Significant gap formation in the modified Bunnell pullout group is of concern although this needs to be correlated in the clinical setting. The lower failure rate of the micro-Mitek group may leave a narrow margin of safety for passive rehabilitation.