CYCLICAL TESTING OF THE MICRO-MITEK SUTURE ANCHOR USED FOR FDP TENDON FIXATION

Eddy Dona, Kim Latendresse, Peter James Scougall, Mark C. Gillies, William Robert Walsh

Open publisher page 0 citations

Abstract

The behaviour of two different methods of reattachment of the flexor digitorum profundus tendon insertion was assessed. Cyclical testing simulating the first 5 days of a passive mobilisation protocol was used to compare the micro Mitek anchor to the modified-Bunnell pull-out suture. Twelve fresh-frozen cadaveric fingers were dissected to the insertion of the FDP tendon. The FDP insertion was then sharply dissected from the distal phalanx and repaired using one of two methods: group 1 -modified Bunnell pullout suture using 3/0 Prolene; group 2 micro Mitek anchor loaded with 3/0 Ethibond inserted into the distal phalanx. Each repaired finger was mounted on to a material testing machine using pneumatic clamps. We cyclically tested the repair between 2N and 15N using a load control of 5N/s for a total of 500 cycles. Gap formation at the tendon bone interface was measured every 100 cycles. No specimens failed during cyclical testing. After 500 cycles, gap formation of the tendon-bone interface was 6.6mm (SD = 1.2mm), and 2.1 mm (SD = 0.3mm) for the pullout technique and the micro Mitek anchor repair respectively. Concerns related to suture anchors, such as anchor failure or protrusion, joint penetration, and anchor-suture junction failure, were not encountered in this study. Cyclical loading results suggest that the repair achieved with both methods of fixation is sufficient to avoid failure. However, significant gap formation at the tendon-bone interface in the modified Bunnell group is of concern, suggesting it may not be the ideal fixation method.

About this research paper

What this paper is about

The behaviour of two different methods of reattachment of the flexor digitorum profundus tendon insertion was assessed. Cyclical testing simulating the first 5 days of a passive mobilisation protocol was used to compare the micro Mitek anchor to the modified-Bunnell pull-out suture. Twelve fresh-frozen cadaveric fingers were dissected to the insertion of the FDP tendon. The FDP insertion was then sharply dissected from the distal phalanx and repaired using one of two methods: group 1 -modified Bunnell pullout suture using 3/0 Prolene; group 2 micro Mitek anchor loaded with 3/0 Ethibond inserted into the distal phalanx. Each repaired finger was mounted on to a material testing machine using pneumatic clamps. We cyclically tested the repair between 2N and 15N using a load control of 5N/s for a total of 500 cycles. Gap formation at the tendon bone interface was measured every 100 cycles. No specimens failed during cyclical testing. After 500 cycles, gap formation of the tendon-bone interface was 6.6mm (SD = 1.2mm), and 2.1 mm (SD = 0.3mm) for the pullout technique and the micro Mitek anchor repair respectively. Concerns related to suture anchors, such as anchor failure or protrusion, joint penetration, and anchor-suture junction failure, were not encountered in this study. Cyclical loading results suggest that the repair achieved with both methods of fixation is sufficient to avoid failure. However, significant gap formation at the tendon-bone interface in the modified Bunnell group is of concern, suggesting it may not be the ideal fixation method.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

The behaviour of two different methods of reattachment of the flexor digitorum profundus tendon insertion was assessed. Cyclical testing simulating the first 5 days of a passive mobilisation protocol was used to compare the micro Mitek anchor to the modified-Bunnell pull-out suture. Twelve fresh-frozen cadaveric fingers were dissected to the insertion of the FDP tendon. The FDP insertion was then sharply dissected from the distal phalanx and repaired using one of two methods: group 1 -modified Bunnell pullout suture using 3/0 Prolene; group 2 micro Mitek anchor loaded with 3/0 Ethibond inserted into the distal phalanx. Each repaired finger was mounted on to a material testing machine using pneumatic clamps. We cyclically tested the repair between 2N and 15N using a load control of 5N/s for a total of 500 cycles. Gap formation at the tendon bone interface was measured every 100 cycles. No specimens failed during cyclical testing. After 500 cycles, gap formation of the tendon-bone interface was 6.6mm (SD = 1.2mm), and 2.1 mm (SD = 0.3mm) for the pullout technique and the micro Mitek anchor repair respectively. Concerns related to suture anchors, such as anchor failure or protrusion, joint penetration, and anchor-suture junction failure, were not encountered in this study. Cyclical loading results suggest that the repair achieved with both methods of fixation is sufficient to avoid failure. However, significant gap formation at the tendon-bone interface in the modified Bunnell group is of concern, suggesting it may not be the ideal fixation method.

Key concepts: Cadaveric spasm, Tendon, Phalanx, Fixation (population genetics), Medicine, Fibrous joint, Prolene, Surgery

Related papers

Back to paper searchBrowse research topicsOriginal source
CYCLICAL TESTING OF THE MICRO-MITEK SUTURE ANCHOR USED FOR FDP TENDON FIXATION — Research Paper | ScholarLens