Comparative analysis between OLYMPUS AU5400 and Hitachi 7170 Auto-biochemical Analyzers

Zhao Lu-wu

Open publisher page 0 citations

Abstract

Objective To investigate the possibility of results comparison made by different biochemical inspection devices and the possibility of its clinic acceptability.Methods According to EP9-A document request of National Committee for Clinical Laboratory of Standards,the comparative method(Y) was AU5400,Hitachi 7170 automatic biochemical analyzer was examination method(X).The co relativities of each substance were analyzed respectively.Results Within day and day to day,coefficients of variation were 0.40%-2.62% and 0.54%-3.33% respectively.Compared with that of OLYMPUS AU5400,the correlation coefficient was between 0.995-0.999.Conclusion The comparability of results in two detecting systems accorded to evaluation of the clinical is acceptable.

About this research paper

What this paper is about

Objective To investigate the possibility of results comparison made by different biochemical inspection devices and the possibility of its clinic acceptability.Methods According to EP9-A document request of National Committee for Clinical Laboratory of Standards,the comparative method(Y) was AU5400,Hitachi 7170 automatic biochemical analyzer was examination method(X).The co relativities of each substance were analyzed respectively.Results Within day and day to day,coefficients of variation were 0.40%-2.62% and 0.54%-3.33% respectively.Compared with that of OLYMPUS AU5400,the correlation coefficient was between 0.995-0.999.Conclusion The comparability of results in two detecting systems accorded to evaluation of the clinical is acceptable.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To investigate the possibility of results comparison made by different biochemical inspection devices and the possibility of its clinic acceptability.Methods According to EP9-A document request of National Committee for Clinical Laboratory of Standards,the comparative method(Y) was AU5400,Hitachi 7170 automatic biochemical analyzer was examination method(X).The co relativities of each substance were analyzed respectively.Results Within day and day to day,coefficients of variation were 0.40%-2.62% and 0.54%-3.33% respectively.Compared with that of OLYMPUS AU5400,the correlation coefficient was between 0.995-0.999.Conclusion The comparability of results in two detecting systems accorded to evaluation of the clinical is acceptable.

Key concepts: Spectrum analyzer, Comparability, Coefficient of variation, Medicine, Correlation coefficient, Medical physics, Mathematics, Statistics

Related papers

Back to paper searchBrowse research topicsOriginal source
Comparative analysis between OLYMPUS AU5400 and Hitachi 7170 Auto-biochemical Analyzers — Research Paper | ScholarLens