Comparative Analysis between OLYMPUS AU5400 and VITROS5.1FS Auto-biochemical Analyzer
Jun Li
Abstract
Jun Li
Abstract
Objective:To investigate the possibility of results comparision made by different biochemical inspection devices and the possibility of its clinic acceptability.Methods:According to EP9-A document request of National Committee for Clinical Laboratory of Standards(NCCLS),the comparative method(Y) was AU5400,Vitros5.1 FS was examination method(X).The linearities,recoveries,and correlativities of each substance were analyzed respectively.Results:Within-day and day-to-day,coefficients of variation were 0.40%~2.62% and 0.54%~3.33% respectively.The dilution degree of samples and the concentration values determined had good linear relationship.Compared with that of OLYMPUS AU5400,the correlation coefficient was between 0.910~0.997.Conclusion:The comparability of results in two detecting systems accorded to evaluation of the clinical is acceptable.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective:To investigate the possibility of results comparision made by different biochemical inspection devices and the possibility of its clinic acceptability.Methods:According to EP9-A document request of National Committee for Clinical Laboratory of Standards(NCCLS),the comparative method(Y) was AU5400,Vitros5.1 FS was examination method(X).The linearities,recoveries,and correlativities of each substance were analyzed respectively.Results:Within-day and day-to-day,coefficients of variation were 0.40%~2.62% and 0.54%~3.33% respectively.The dilution degree of samples and the concentration values determined had good linear relationship.Compared with that of OLYMPUS AU5400,the correlation coefficient was between 0.910~0.997.Conclusion:The comparability of results in two detecting systems accorded to evaluation of the clinical is acceptable.
Key concepts: Comparability, Coefficient of variation, Correlation coefficient, Spectrum analyzer, Linear correlation, Statistics, Mathematics, Chromatography