Simultaneous Determination of Arsenic and Mercury in Purified Water by HG-ICP-AES
Ying Liu
Abstract
Ying Liu
Abstract
A HG\|ICP\|AES method for simultaneous determination of arsenic and mercury in purified water is proposed.The detection limit of arsenic and mercury is 1.7μg/L and 0.1μg/L respectively.The RSD for arsenic is 3.46% and for mercury is 2.85%.The recovery for arsenic is in the range of 92%—110% and for mercury is in the range of 90%—113%.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
A HG\|ICP\|AES method for simultaneous determination of arsenic and mercury in purified water is proposed.The detection limit of arsenic and mercury is 1.7μg/L and 0.1μg/L respectively.The RSD for arsenic is 3.46% and for mercury is 2.85%.The recovery for arsenic is in the range of 92%—110% and for mercury is in the range of 90%—113%.
Key concepts: Mercury (programming language), Arsenic, Chemistry, Environmental chemistry, Detection limit, Radiochemistry, Chromatography, Computer science