Comparative analysis and bias assessment on results of partial biochemical items detected by different biochemical testing systems
Zuowei Yuan
Abstract
Zuowei Yuan
Abstract
Objective:To investigate the comparability among different biochemical testing systems in the same laboratory by performing the comparative analysis and bias estimation on the results.Methods:According to NCCLS document EP9-A2,3 biochemical testing systems including system 2 and 3(Y1-Y2,comparative testing systems) and system 1(X,targeted testing system,which was composed of Hitachi 7600 biochemical analyzer P1,maker reagents,c.f.a.s.calibrator and BIO RAD quality control) were used to obtain correlation coefficient and linear regression equation.The values of aspartate aminotransferase(AST),alamine aminotransferase(ALT),gamma-glutamyl transferase(GGT),alkaline phosphatase(ALP),Urea,creatinine(Cr) and uric acid(UA) in patient's serum were detected by the three systems.Then the coefficient and the equation were used to evaluate the system bias between comparative testing systems and targeted testing system.The comparability of different testing systems was judged by half of the CLIA'88 standard.Results:The results of the three testing systems were significantly correlated(r20.95).The results of all testing items showed good comparability except those of Urea(with low concentration level) tested by system 2(Y2) and system 1(X)(%SE1/2CILA'88 TEa).Conclusions:There is good comparability among the results of AST、ALT、GGT、ALP、Cr、UA detected by different testing systems in our hospital except those of Urea with low concentration level.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective:To investigate the comparability among different biochemical testing systems in the same laboratory by performing the comparative analysis and bias estimation on the results.Methods:According to NCCLS document EP9-A2,3 biochemical testing systems including system 2 and 3(Y1-Y2,comparative testing systems) and system 1(X,targeted testing system,which was composed of Hitachi 7600 biochemical analyzer P1,maker reagents,c.f.a.s.calibrator and BIO RAD quality control) were used to obtain correlation coefficient and linear regression equation.The values of aspartate aminotransferase(AST),alamine aminotransferase(ALT),gamma-glutamyl transferase(GGT),alkaline phosphatase(ALP),Urea,creatinine(Cr) and uric acid(UA) in patient's serum were detected by the three systems.Then the coefficient and the equation were used to evaluate the system bias between comparative testing systems and targeted testing system.The comparability of different testing systems was judged by half of the CLIA'88 standard.Results:The results of the three testing systems were significantly correlated(r20.95).The results of all testing items showed good comparability except those of Urea(with low concentration level) tested by system 2(Y2) and system 1(X)(%SE1/2CILA'88 TEa).Conclusions:There is good comparability among the results of AST、ALT、GGT、ALP、Cr、UA detected by different testing systems in our hospital except those of Urea with low concentration level.
Key concepts: Comparability, Uric acid, Creatinine, Alanine aminotransferase, Spectrum analyzer, Alkaline phosphatase, Urea, Mathematics