Contrastive Experiments of automatic biochemical analyzer on glucose
Han Bao-lian
Abstract
Han Bao-lian
Abstract
Objective:Comparing the glucose test results of Siemens Dimension Xpand automatic biochemical analyzer and domestic Dirui CS-600 automatic biochemical analyzer according to the NCCLS EP9-A file,to explore the comparability for blood glucose test of the two instruments.Methods:We took Diri CS-600 automatic biochemical analyzer as reference instrument,and Siemens Dimension Xpand automatic biochemical analyzer as a test one.Fresh serum selected were tested by the two types of instruments everyday,obtain the correlation coefficient and regression equation of Y=bx+a by the statistical software from the test results,with USA amendment clinical laboratory(CLLA,88) provisions of 1/2 EQA(external quality assessment) permissible error as the standard,to judge the comparability of test results of the two instruments.Results:Differences between the two instruments test of blood glucose was not statistically significant(r = 0.9997,P 0.975).Conclusion:Blood glucose test results of the two instruments have good comparability.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective:Comparing the glucose test results of Siemens Dimension Xpand automatic biochemical analyzer and domestic Dirui CS-600 automatic biochemical analyzer according to the NCCLS EP9-A file,to explore the comparability for blood glucose test of the two instruments.Methods:We took Diri CS-600 automatic biochemical analyzer as reference instrument,and Siemens Dimension Xpand automatic biochemical analyzer as a test one.Fresh serum selected were tested by the two types of instruments everyday,obtain the correlation coefficient and regression equation of Y=bx+a by the statistical software from the test results,with USA amendment clinical laboratory(CLLA,88) provisions of 1/2 EQA(external quality assessment) permissible error as the standard,to judge the comparability of test results of the two instruments.Results:Differences between the two instruments test of blood glucose was not statistically significant(r = 0.9997,P 0.975).Conclusion:Blood glucose test results of the two instruments have good comparability.
Key concepts: Spectrum analyzer, Comparability, Medicine, Siemens, Statistics, Medical physics, Mathematics, Computer science