Comparison of plasma glucose results tested in two different automatic biochemical analyzers

Zhiwei Wang

Open publisher page 0 citations

Abstract

Objective To assess the comparability of measurement results for plasma glucose from two automatic biochemical systems.Methods According to the requirements in American EP9-A2 file, plasma glucose concentrations in the same plasma samples were determined, using ROCHE P800 as a standard system and TOSIBA ACCTUE as the experimental one. The clinical acceptability of the two kinds of biochemical analyzers was judged based on the criteria for quality assessment among clinical laboratories in American amend legislation (CLIA'88)in which the required criteria is defined as one third of allowable error.Results The glucose results from two kinds of instruments showed a good correlation(r0.975); clinical acceptability of the instruments was in the acceptable range.Conclusion The detected glucose results of two instruments have good comparability.

About this research paper

What this paper is about

Objective To assess the comparability of measurement results for plasma glucose from two automatic biochemical systems.Methods According to the requirements in American EP9-A2 file, plasma glucose concentrations in the same plasma samples were determined, using ROCHE P800 as a standard system and TOSIBA ACCTUE as the experimental one. The clinical acceptability of the two kinds of biochemical analyzers was judged based on the criteria for quality assessment among clinical laboratories in American amend legislation (CLIA'88)in which the required criteria is defined as one third of allowable error.Results The glucose results from two kinds of instruments showed a good correlation(r0.975); clinical acceptability of the instruments was in the acceptable range.Conclusion The detected glucose results of two instruments have good comparability.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To assess the comparability of measurement results for plasma glucose from two automatic biochemical systems.Methods According to the requirements in American EP9-A2 file, plasma glucose concentrations in the same plasma samples were determined, using ROCHE P800 as a standard system and TOSIBA ACCTUE as the experimental one. The clinical acceptability of the two kinds of biochemical analyzers was judged based on the criteria for quality assessment among clinical laboratories in American amend legislation (CLIA'88)in which the required criteria is defined as one third of allowable error.Results The glucose results from two kinds of instruments showed a good correlation(r0.975); clinical acceptability of the instruments was in the acceptable range.Conclusion The detected glucose results of two instruments have good comparability.

Key concepts: Comparability, Medicine, Plasma glucose, Medical physics, Statistics, Internal medicine, Mathematics, Insulin

Related papers

Back to paper searchBrowse research topicsOriginal source
Comparison of plasma glucose results tested in two different automatic biochemical analyzers — Research Paper | ScholarLens