Comparative study about CMIA and two kinds of ELISA on low level of HBsAg samples
Pei Lu
Abstract
Pei Lu
Abstract
Objective To compare the sensitivity between China-made one-step hepatitis B surface antigen(HBsAg) enzyme linked immunosorbent assay(ELISA) reagent and two-step HBsAg ELISA reagent supplied by same company.Methods HBsAg samples were detected by one-step ELISA and two-step ELISA,at the same time compared with the result tested by CMIA.Results In 90 cases,the results of one-step ELISA and two-step ELISA were all negative while CMIA were negative.Among 61 low level samples detected by CMIA,15 positive samples were detected by one-step ELISA,46 positive samples were detected by two-step ELISA.The sensitivity of two-step ELISA was significantly higher than that of one-step ELISA(P0.01).One-step ELISA had the same sensitivity of detection with CMIA above 2.50 U/mL.Two-step ELISA had the same sensitivity of detection with CMIA above 0.50 U/mL.The sensitivity of two-step ELISA was significantly lower than CMIA under 0.50 U/mL(P0.01).Conclusion The sensitivity of two-step ELISA is significantly higher than one-step ELISA,but is still less than CMIA in the lower level samples.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To compare the sensitivity between China-made one-step hepatitis B surface antigen(HBsAg) enzyme linked immunosorbent assay(ELISA) reagent and two-step HBsAg ELISA reagent supplied by same company.Methods HBsAg samples were detected by one-step ELISA and two-step ELISA,at the same time compared with the result tested by CMIA.Results In 90 cases,the results of one-step ELISA and two-step ELISA were all negative while CMIA were negative.Among 61 low level samples detected by CMIA,15 positive samples were detected by one-step ELISA,46 positive samples were detected by two-step ELISA.The sensitivity of two-step ELISA was significantly higher than that of one-step ELISA(P0.01).One-step ELISA had the same sensitivity of detection with CMIA above 2.50 U/mL.Two-step ELISA had the same sensitivity of detection with CMIA above 0.50 U/mL.The sensitivity of two-step ELISA was significantly lower than CMIA under 0.50 U/mL(P0.01).Conclusion The sensitivity of two-step ELISA is significantly higher than one-step ELISA,but is still less than CMIA in the lower level samples.
Key concepts: HBsAg, Two step, Reagent, Medicine, Chromatography, Virology, Chemistry, Hepatitis B virus