Calibration and comparative analysis among different hematology analyzers
Gao Hong-chen
Abstract
Gao Hong-chen
Abstract
Objective To improve the accuracy and consistency,we compared the test results from the different blood cell analyzers to explore whether the variance of the results were within the permitted range.Method The one hematology analyzer with best performance was chosen as reference analyzer,anticoagulative fresh whole blood was detected on several hematology analyzers by verification test,then we evaluated test results with the reference provided by SYSMEX company.Results The test results are all in reference limits with good accuracy and consistency except one WBC value of XS-800i.Conclusion When the laboratory has more than one blood cell analyzer,we should regularly perform the verifying test and the calibrating with fresh whole blood to establish calibration and quality control methods of the multiple blood cell analyzers in order to ensure the accuracy of laboratory test results and consistency.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To improve the accuracy and consistency,we compared the test results from the different blood cell analyzers to explore whether the variance of the results were within the permitted range.Method The one hematology analyzer with best performance was chosen as reference analyzer,anticoagulative fresh whole blood was detected on several hematology analyzers by verification test,then we evaluated test results with the reference provided by SYSMEX company.Results The test results are all in reference limits with good accuracy and consistency except one WBC value of XS-800i.Conclusion When the laboratory has more than one blood cell analyzer,we should regularly perform the verifying test and the calibrating with fresh whole blood to establish calibration and quality control methods of the multiple blood cell analyzers in order to ensure the accuracy of laboratory test results and consistency.
Key concepts: Hematology analyzer, Medicine, Calibration, Spectrum analyzer, Consistency (knowledge bases), Reference range, Hematology, Test (biology)