2011Modern Medicine HealthRequires access

Study of comparison and calibration of different hematology analyzers using fresh blood sample

Fangjiu Liu

Open publisher page 0 citations

Abstract

Objective:To evaluate comparability and the accuracy among 8 different hematology analyzers in our hospital.Methods:The precision of 8 different hematology analyzers was determined by the repeatability test.The comparative test was performed with fresh blood sample.The definite values of fresh blood were determined with reference hematology analyzer,and the rest 7 analyzers were calibrated with the definite value.Results:The errors existed among different hematology analyzers.So calibration was necessary.The 8 analyzers had good consistency and accuracy after calibration with allowable range of errors.Conclusion:The method of calibration with fresh blood sample is simple,convenient,reliable and low-cost,which is worthy of being promoted.

About this research paper

What this paper is about

Objective:To evaluate comparability and the accuracy among 8 different hematology analyzers in our hospital.Methods:The precision of 8 different hematology analyzers was determined by the repeatability test.The comparative test was performed with fresh blood sample.The definite values of fresh blood were determined with reference hematology analyzer,and the rest 7 analyzers were calibrated with the definite value.Results:The errors existed among different hematology analyzers.So calibration was necessary.The 8 analyzers had good consistency and accuracy after calibration with allowable range of errors.Conclusion:The method of calibration with fresh blood sample is simple,convenient,reliable and low-cost,which is worthy of being promoted.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective:To evaluate comparability and the accuracy among 8 different hematology analyzers in our hospital.Methods:The precision of 8 different hematology analyzers was determined by the repeatability test.The comparative test was performed with fresh blood sample.The definite values of fresh blood were determined with reference hematology analyzer,and the rest 7 analyzers were calibrated with the definite value.Results:The errors existed among different hematology analyzers.So calibration was necessary.The 8 analyzers had good consistency and accuracy after calibration with allowable range of errors.Conclusion:The method of calibration with fresh blood sample is simple,convenient,reliable and low-cost,which is worthy of being promoted.

Key concepts: Hematology analyzer, Medicine, Repeatability, Calibration, Hematology, Sample (material), Reference range, Internal medicine

Related papers

Back to paper searchBrowse research topicsOriginal source
Study of comparison and calibration of different hematology analyzers using fresh blood sample — Research Paper | ScholarLens