The Influence of Anticoagulants and Time of Measurements on D-Dimer Determinations
Hi-Jeong Kwon
Abstract
Hi-Jeong Kwon
Abstract
Background : We evaluated the effects of anticoagulants used in blood sampling and time of mesurements after blood collection on D-dimer determinations. Methods : A total of 32 samples of sodium citrated plasma, and 23 samples each of serum and ethylene diamine tetraacetic acid (EDTA) plasma, were studied. D-dimer was measured by enzyme linked fluorescent assay using a VIDAS analyzer. Results : The meanSD (median) of D-dimer titers in citrated plasma, serum, and EDTA plasma samples immediately after blood collection were 282.7201.1 (224.6)g/L, 255.1206.8 (197.3) g/L, and 313.3230.8 (246.4)g/L respectively. The results obtained for the three different sample groups were well correlated. While the upper limit of the reference range for D-dimer for citrated plasma has been set at 500g/L, the cut-off values for serum and EDTA plasma were 400g/L (100% sensitivity and specificity), and 585g/L (100% sensitivity and 94.7% specificity), respectively Conclusions : In addition to citrated plasma, both serum and EDTA plasma can be used for D-dimer assay. A delay in the time of measurements after blood collection does not influence the results significantly.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Background : We evaluated the effects of anticoagulants used in blood sampling and time of mesurements after blood collection on D-dimer determinations. Methods : A total of 32 samples of sodium citrated plasma, and 23 samples each of serum and ethylene diamine tetraacetic acid (EDTA) plasma, were studied. D-dimer was measured by enzyme linked fluorescent assay using a VIDAS analyzer. Results : The meanSD (median) of D-dimer titers in citrated plasma, serum, and EDTA plasma samples immediately after blood collection were 282.7201.1 (224.6)g/L, 255.1206.8 (197.3) g/L, and 313.3230.8 (246.4)g/L respectively. The results obtained for the three different sample groups were well correlated. While the upper limit of the reference range for D-dimer for citrated plasma has been set at 500g/L, the cut-off values for serum and EDTA plasma were 400g/L (100% sensitivity and specificity), and 585g/L (100% sensitivity and 94.7% specificity), respectively Conclusions : In addition to citrated plasma, both serum and EDTA plasma can be used for D-dimer assay. A delay in the time of measurements after blood collection does not influence the results significantly.
Key concepts: D-dimer, Chemistry, Dimer, Reference range, Chromatography, Ethylene diamine, Titer, Blood sampling