Verification Assessment in Three Hematology Analyzers

Xinming Liu

Open publisher page 0 citations

Abstract

Objective To evaluate the veracity and consistency of the results obtained from three automated hematology an-alyzers. Methods The reproducibility tests were used to assess the precision,the verification tests were exerted to evaluate the accuracy and consistency of test parameters(WBC,RBC,Hb,Hct and Plt) of the three hematology analyzers. According to the 1/2 standard of American Clinical Laboratory Improvement Amendment (CLIA,88),the relative bias of the reference analyzer and other assessed analyzers were determined. Results The results obtained from the three analyzers showed ex-cellent precision and stability,the coefficients of variation (CV) were all less than 5%; compared with the results of LH750,the results of Gens 5diff and ACT-5diff showed excellent correlation,all coefficients of correlation(γ) exceeded 0.975. The results of both RBC and Hct obtained from Gens 5diff and ACT-5diff exceeded allowable error; the results of WBC,Hb and Plt were in the allowable error range. Conclusion In the condition of the excellent precision and stability,system errors existed in the analyzer can be found by verification on time. Through being adjusted and calibrated,the hematology ana-lyzers can meet the requirements and these results are consistent.

About this research paper

What this paper is about

Objective To evaluate the veracity and consistency of the results obtained from three automated hematology an-alyzers. Methods The reproducibility tests were used to assess the precision,the verification tests were exerted to evaluate the accuracy and consistency of test parameters(WBC,RBC,Hb,Hct and Plt) of the three hematology analyzers. According to the 1/2 standard of American Clinical Laboratory Improvement Amendment (CLIA,88),the relative bias of the reference analyzer and other assessed analyzers were determined. Results The results obtained from the three analyzers showed ex-cellent precision and stability,the coefficients of variation (CV) were all less than 5%; compared with the results of LH750,the results of Gens 5diff and ACT-5diff showed excellent correlation,all coefficients of correlation(γ) exceeded 0.975. The results of both RBC and Hct obtained from Gens 5diff and ACT-5diff exceeded allowable error; the results of WBC,Hb and Plt were in the allowable error range. Conclusion In the condition of the excellent precision and stability,system errors existed in the analyzer can be found by verification on time. Through being adjusted and calibrated,the hematology ana-lyzers can meet the requirements and these results are consistent.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To evaluate the veracity and consistency of the results obtained from three automated hematology an-alyzers. Methods The reproducibility tests were used to assess the precision,the verification tests were exerted to evaluate the accuracy and consistency of test parameters(WBC,RBC,Hb,Hct and Plt) of the three hematology analyzers. According to the 1/2 standard of American Clinical Laboratory Improvement Amendment (CLIA,88),the relative bias of the reference analyzer and other assessed analyzers were determined. Results The results obtained from the three analyzers showed ex-cellent precision and stability,the coefficients of variation (CV) were all less than 5%; compared with the results of LH750,the results of Gens 5diff and ACT-5diff showed excellent correlation,all coefficients of correlation(γ) exceeded 0.975. The results of both RBC and Hct obtained from Gens 5diff and ACT-5diff exceeded allowable error; the results of WBC,Hb and Plt were in the allowable error range. Conclusion In the condition of the excellent precision and stability,system errors existed in the analyzer can be found by verification on time. Through being adjusted and calibrated,the hematology ana-lyzers can meet the requirements and these results are consistent.

Key concepts: Hematology analyzer, Reproducibility, Consistency (knowledge bases), Spectrum analyzer, Medicine, Repeatability, Coefficient of variation, Hematology