Comparability Investigation for Results of aPTT and TT Between Two Difference Blood Coagulation Detection Systems
Tang Jin-guang
Abstract
Tang Jin-guang
Abstract
Objective To improve the comparability for results of activated partial tromboplastin time (aPTT) and thrombin time (TT) from two difference blood coagulation detection systems.Methods Fresh plasma were measured by a calibrated reference ACL-Advance blood coagulation analyzer.Two plasma with 30 seconds and 12.5 seconds in aPTT and TT measurement were selected as standardized samples to adjust the results of detection by the Sysmex CA-50 blood coagulation analyzer.The results were compared in low, middle and high level of aPTT and TT measurement from two analyzers respectively.Results There was no significant difference in results of low,middle and high level of aPTT and TT measurement between two adjus- ted blood coagulation analyzers (P>0.05).Excellent correlation was found between these two analyzers (r(?) 0.98).Conclusion Comparability can be improved between two blood coagulation detection systems by using the same reagent and the standardized fresh plasma measured previously by a calibrated reference analy- zer to adjust another one.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To improve the comparability for results of activated partial tromboplastin time (aPTT) and thrombin time (TT) from two difference blood coagulation detection systems.Methods Fresh plasma were measured by a calibrated reference ACL-Advance blood coagulation analyzer.Two plasma with 30 seconds and 12.5 seconds in aPTT and TT measurement were selected as standardized samples to adjust the results of detection by the Sysmex CA-50 blood coagulation analyzer.The results were compared in low, middle and high level of aPTT and TT measurement from two analyzers respectively.Results There was no significant difference in results of low,middle and high level of aPTT and TT measurement between two adjus- ted blood coagulation analyzers (P>0.05).Excellent correlation was found between these two analyzers (r(?) 0.98).Conclusion Comparability can be improved between two blood coagulation detection systems by using the same reagent and the standardized fresh plasma measured previously by a calibrated reference analy- zer to adjust another one.
Key concepts: Medicine, Coagulation, Prothrombin time, Significant difference, Coagulation testing, Spectrum analyzer, Comparability, Coagulation system