Determination of Selenium in Water Sample by Hydride Generation-Atomic Fluorescence Spectrometry with Microwave Digestion
Han Guo-ca
Abstract
Han Guo-ca
Abstract
Objective To establish a method for detection of selenium in water sample. Methods The method of digestion with nitric acid-hydrogen peroxide was used. Results The calibration curve for selenium was 0-100 μg/L, r=0.999 2, detection limit 0.42 μg/L and the recovery of standard addition 98.9% to 102.3%. Conclusion This method is simple, rapid, sensitive and of accurate, the method has been applied to the sample analysis with satisfactory results.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To establish a method for detection of selenium in water sample. Methods The method of digestion with nitric acid-hydrogen peroxide was used. Results The calibration curve for selenium was 0-100 μg/L, r=0.999 2, detection limit 0.42 μg/L and the recovery of standard addition 98.9% to 102.3%. Conclusion This method is simple, rapid, sensitive and of accurate, the method has been applied to the sample analysis with satisfactory results.