Performance evaluation of Mairui BS-300 automatic biochemical analyzer

Wang Hui-min

Open publisher page 0 citations

Abstract

Objective: To evaluate the main characteristics of BS-300 automatic biochemistry analyzer in our laboratory.Methods: Evaluating the accuracy,repeatability and linearity of absorbance,stray light,cross contamination by standard solutions offered by the centre of national standard substances.Examining the absorbance difference between cuvettes by potassium dichromate solution.Observing intra imprecision by patients' serum.Results: The mean absorbance of standard material 1-1 and 1-2 were 0.407 and 0.912,respectively,the repeatability of absorbance was 0.008 4.The linearity errors of standard solutions 3-1,3-2,3-3,3-4 and 3-5 were-4.15%,0.46%,2.60%,0.67%,0.42%,respectively.The stray light of standard solution 2 was 6.880 6.The probability of cross-contamination of standard solution 4 were-1.58% and-0.26%,respectively.The CV of the absorbance between cuvettes was 1.35%.The CV of ALP at low level was 1.14%,and at high level was 0.78%.Conclusion: According to the document of JJG464-2005,the absorbance difference between cuvettes,intra imprecision,stray light and linearity errors meet the requirements.But the accuracy,repeatability and the probability of cross contamination can not meet the requirements.

About this research paper

What this paper is about

Objective: To evaluate the main characteristics of BS-300 automatic biochemistry analyzer in our laboratory.Methods: Evaluating the accuracy,repeatability and linearity of absorbance,stray light,cross contamination by standard solutions offered by the centre of national standard substances.Examining the absorbance difference between cuvettes by potassium dichromate solution.Observing intra imprecision by patients' serum.Results: The mean absorbance of standard material 1-1 and 1-2 were 0.407 and 0.912,respectively,the repeatability of absorbance was 0.008 4.The linearity errors of standard solutions 3-1,3-2,3-3,3-4 and 3-5 were-4.15%,0.46%,2.60%,0.67%,0.42%,respectively.The stray light of standard solution 2 was 6.880 6.The probability of cross-contamination of standard solution 4 were-1.58% and-0.26%,respectively.The CV of the absorbance between cuvettes was 1.35%.The CV of ALP at low level was 1.14%,and at high level was 0.78%.Conclusion: According to the document of JJG464-2005,the absorbance difference between cuvettes,intra imprecision,stray light and linearity errors meet the requirements.But the accuracy,repeatability and the probability of cross contamination can not meet the requirements.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective: To evaluate the main characteristics of BS-300 automatic biochemistry analyzer in our laboratory.Methods: Evaluating the accuracy,repeatability and linearity of absorbance,stray light,cross contamination by standard solutions offered by the centre of national standard substances.Examining the absorbance difference between cuvettes by potassium dichromate solution.Observing intra imprecision by patients' serum.Results: The mean absorbance of standard material 1-1 and 1-2 were 0.407 and 0.912,respectively,the repeatability of absorbance was 0.008 4.The linearity errors of standard solutions 3-1,3-2,3-3,3-4 and 3-5 were-4.15%,0.46%,2.60%,0.67%,0.42%,respectively.The stray light of standard solution 2 was 6.880 6.The probability of cross-contamination of standard solution 4 were-1.58% and-0.26%,respectively.The CV of the absorbance between cuvettes was 1.35%.The CV of ALP at low level was 1.14%,and at high level was 0.78%.Conclusion: According to the document of JJG464-2005,the absorbance difference between cuvettes,intra imprecision,stray light and linearity errors meet the requirements.But the accuracy,repeatability and the probability of cross contamination can not meet the requirements.

Key concepts: Repeatability, Absorbance, Cuvette, Potassium dichromate, Linearity, Stray light, Standard solution, Spectrum analyzer