Comparative analysis of detection results for homocysteine levels by two chemical luminescence immunity analyzers

Lu Mi

Open publisher page 0 citations

Abstract

Objective To explore the comparability of the results between different instruments in the same laboratory. Methods Forty patients were included. The HCY in serum was detected by ABBOTT I2000 and ABBOTT Axsym separately and the results were studied by correlation coefficient analysis. Results The results showed that the correlation coefficient(r) was 0.9998 and its regression equation was y=0.9846x-0.0443. The expected relative deviations were 1.7% and 1.6% respectively at medical decision level. There was closely correlation between the results from the two different machines. Conclusion When different instruments are used to detect the same project in one lab, the correlation analysis and bias evaluation should be carried out to ensure the comparability and consistency of the different instruments.

About this research paper

What this paper is about

Objective To explore the comparability of the results between different instruments in the same laboratory. Methods Forty patients were included. The HCY in serum was detected by ABBOTT I2000 and ABBOTT Axsym separately and the results were studied by correlation coefficient analysis. Results The results showed that the correlation coefficient(r) was 0.9998 and its regression equation was y=0.9846x-0.0443. The expected relative deviations were 1.7% and 1.6% respectively at medical decision level. There was closely correlation between the results from the two different machines. Conclusion When different instruments are used to detect the same project in one lab, the correlation analysis and bias evaluation should be carried out to ensure the comparability and consistency of the different instruments.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To explore the comparability of the results between different instruments in the same laboratory. Methods Forty patients were included. The HCY in serum was detected by ABBOTT I2000 and ABBOTT Axsym separately and the results were studied by correlation coefficient analysis. Results The results showed that the correlation coefficient(r) was 0.9998 and its regression equation was y=0.9846x-0.0443. The expected relative deviations were 1.7% and 1.6% respectively at medical decision level. There was closely correlation between the results from the two different machines. Conclusion When different instruments are used to detect the same project in one lab, the correlation analysis and bias evaluation should be carried out to ensure the comparability and consistency of the different instruments.

Key concepts: Medicine, Comparability, Correlation coefficient, Statistics, Correlation, Consistency (knowledge bases), Internal medicine, Medical physics