Determination of Palladium in Fresh and Spent Pd/C Catalyst by ICP-AES
Fang Haiyan
Abstract
Fang Haiyan
Abstract
The content of Pd was determined through(1) wet digestion with HNO3-HClO4or ignition at 700℃ to remove carbon,(2) dissolution of the residues by aqua regia and(3) determination of palladium by ICP-AES.Pretreatment of sample,accuracy and precision of the method were discussed.The additional standard recovery is 99%~101%,RSD 1%.The method is simple,accurate and fast.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
The content of Pd was determined through(1) wet digestion with HNO3-HClO4or ignition at 700℃ to remove carbon,(2) dissolution of the residues by aqua regia and(3) determination of palladium by ICP-AES.Pretreatment of sample,accuracy and precision of the method were discussed.The additional standard recovery is 99%~101%,RSD 1%.The method is simple,accurate and fast.
Key concepts: Aqua regia, Palladium, Inductively coupled plasma atomic emission spectroscopy, Dissolution, Chemistry, Catalysis, Nuclear chemistry, Metallurgy