ICP-MS Determination of Platinum,Palladium and Rhodium in Vehicle-Use Catalyst with Microwave Assisted Sample Digestion

Zhou Hai-ming

Open publisher page 0 citations

Abstract

Sample of vehicle-use catalysts was digested with a mixture of 6 mL HCl,3 mL HNO3,1 mL HF and 4 mL H2O,and an aliquot of the sample solution was taken for ICP-MS determination of Pt,Pd and Rh.195Pt,105Pd and 103Rh were used as measuring elements.Values of detection limit(3σ) of Pt,Pd and Rh found were 0.003,0.5,0.002 μg·L-1 respectively.The proposed method was used in the analysis of 5 catalyst samples,the results obtained were in conformity with those obtained by 2 other labs.Values of recovery were found in the range of 85.5%-101%.

About this research paper

What this paper is about

Sample of vehicle-use catalysts was digested with a mixture of 6 mL HCl,3 mL HNO3,1 mL HF and 4 mL H2O,and an aliquot of the sample solution was taken for ICP-MS determination of Pt,Pd and Rh.195Pt,105Pd and 103Rh were used as measuring elements.Values of detection limit(3σ) of Pt,Pd and Rh found were 0.003,0.5,0.002 μg·L-1 respectively.The proposed method was used in the analysis of 5 catalyst samples,the results obtained were in conformity with those obtained by 2 other labs.Values of recovery were found in the range of 85.5%-101%.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Sample of vehicle-use catalysts was digested with a mixture of 6 mL HCl,3 mL HNO3,1 mL HF and 4 mL H2O,and an aliquot of the sample solution was taken for ICP-MS determination of Pt,Pd and Rh.195Pt,105Pd and 103Rh were used as measuring elements.Values of detection limit(3σ) of Pt,Pd and Rh found were 0.003,0.5,0.002 μg·L-1 respectively.The proposed method was used in the analysis of 5 catalyst samples,the results obtained were in conformity with those obtained by 2 other labs.Values of recovery were found in the range of 85.5%-101%.

Key concepts: Rhodium, Palladium, Platinum, Detection limit, Catalysis, Inductively coupled plasma mass spectrometry, Microwave digestion, Chemistry

Related papers

Back to paper searchBrowse research topicsOriginal source
ICP-MS Determination of Platinum,Palladium and Rhodium in Vehicle-Use Catalyst with Microwave Assisted Sample Digestion — Research Paper | ScholarLens