The Influence of Central Corneal Thickness of Myopia on the Measurementsof Intraocular Pressure with Non-contact Tonometer
Tao Li
Abstract
Tao Li
Abstract
Objective:To study the relationship between the central corneal thickness(CCT)and the measurements of intraocular pressure(IOP)with non-contact tonometer.Methods:CCT and IOP were mearured on 411 cases who are myopia,the data were analysised by SPSS 10.0.Result:The mean of CCT was 549 μm±31μm,the mean of IOP was 16.74mmHg±2.58mmHg.A positive correlation was found between IOP and CCT(r=0.447,P0.001).Regression equation for IOP(Y)=-7.854+0.0448X(X=CCT,P0.001).According to CCT,all cases were classfied into 3 groups and a significant difference was found between IOP in different groups.Conclusion:The measurements of IOP were influenced significantly by CCT in myopia group.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective:To study the relationship between the central corneal thickness(CCT)and the measurements of intraocular pressure(IOP)with non-contact tonometer.Methods:CCT and IOP were mearured on 411 cases who are myopia,the data were analysised by SPSS 10.0.Result:The mean of CCT was 549 μm±31μm,the mean of IOP was 16.74mmHg±2.58mmHg.A positive correlation was found between IOP and CCT(r=0.447,P0.001).Regression equation for IOP(Y)=-7.854+0.0448X(X=CCT,P0.001).According to CCT,all cases were classfied into 3 groups and a significant difference was found between IOP in different groups.Conclusion:The measurements of IOP were influenced significantly by CCT in myopia group.
Key concepts: Medicine, Ophthalmology, Intraocular pressure, Significant difference, Goldmann Applanation Tonometer, Optometry, Internal medicine