Determination of formate and hydrochloric in air sample by ion chromatography
Xiao Wang
Abstract
Xiao Wang
Abstract
A method for the determination of formate and hydrochloric in air and environmental sample by ion chromatography is described. The result showed that the detection limites of the method for formate and hydrochloric were 0.01mg/L and 0.02mg/L; recoveries were 98%~101% and 98%~102%; the RSDs ware 1.9% and 1.6%; the linrar ranges were 0.00~10mg/L and 0.00~16mg/L. The method is simple, rapid and reliable for determining formate and hydrochloric in rain, exhaust gas, surface water and ground water.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
A method for the determination of formate and hydrochloric in air and environmental sample by ion chromatography is described. The result showed that the detection limites of the method for formate and hydrochloric were 0.01mg/L and 0.02mg/L; recoveries were 98%~101% and 98%~102%; the RSDs ware 1.9% and 1.6%; the linrar ranges were 0.00~10mg/L and 0.00~16mg/L. The method is simple, rapid and reliable for determining formate and hydrochloric in rain, exhaust gas, surface water and ground water.
Key concepts: Formate, Hydrochloric acid, Chemistry, Chromatography, Ion chromatography, Detection limit, Ion, Inorganic chemistry