Quantitative determination of benzetonium by UV spectrophotometry
Ye Si, HE Cai-yu
Abstract
Ye Si, HE Cai-yu
Abstract
Objective To establish a UV spectrophotometry method for the determination of benzetonium and its linear range.Methods A UV method was used to determine the benzetonium at 269nm.Results The concentration and absorption had a good linear relationship in the range of 0.05~0.3 mg/ml.The equation of relatiral regression was A=2.8005xC+0.0332 r=0.9997.The average recovery of benzetonium was100.01%,RSD=0.11%.Conclusion The method is simple and accurate.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To establish a UV spectrophotometry method for the determination of benzetonium and its linear range.Methods A UV method was used to determine the benzetonium at 269nm.Results The concentration and absorption had a good linear relationship in the range of 0.05~0.3 mg/ml.The equation of relatiral regression was A=2.8005xC+0.0332 r=0.9997.The average recovery of benzetonium was100.01%,RSD=0.11%.Conclusion The method is simple and accurate.